The program settings and cleansers and disinfectants that must be
used can be found in the instructions for use of the thermodisinfector.
6.2.3 Cleaning: Manual cleaning of the inside
To effectively set up, the inside of the machine must be cleaned automat‐
ically in a cleaning and disinfection unit in accordance with ISO 15883-1.
(The interior of this product is not to be cleaned manually).
6.2.4 Cleaning: Automated internal cleaning
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g. Miele G 7781/G 7881
(Validation was performed with the program "VARIO-TD", the cleaner "ne‐
odisher
®
mediclean", the neutraliser "neodisher
®
Z" and rinse "neodisher
®
mielclear" and only refers to the material compatibility with KaVo products.)
Preparation methods according to ISO 17664