6.5
Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic Drying
The drying procedure is normally part of the disinfection program of the thermodi‐
sinfector.
Note
Follow the instructions for use of the thermodesinfector (compressed air quality
according to ISO 7494-2).
6.6
Packaging
Note
The sterilisation bag must be large enough for the tip so that the bag is not
stretched.
The quality and use of the sterilised product packaging must satisfy applicable
standards and be suitable for the sterilisation procedure.
Note
The SONICflex can also be sterilised in the sterilisation tray.
▶ Individually weld the medical device in the sterilised item packaging (such as
KaVo STERIclave bags Mat. no. 0.411.9912).
▶ Tips with short thread and marked A only fit in green mounts.
Tips with long thread fit in blue and green mounts.
6.7
Sterilisation
Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisa‐
tion cycle!
The medical device is resistant to temperatures up to 138℃ (280.4°F).
Instructions for use For SONICflex tips paro - REF 0.571.0371, paro A - REF 1.006.2020
6 Preparation methods according to ISO 17664 | 6.5 Drying
15/17