6.3.1
Manual cleaning of the exterior
Accessories required:
▪ Tap water 30 ℃ +/- 5 ℃ (86 °F +/- 10 °F)
▪ Nozzle needle
▪ Rubber eraser
▪ Brush, e.g. medium-hard toothbrush
Brush the SONICflex tip under running tap water using, e.g., a toothbrush of me‐
dium hardness. Clean the cone of the tip with a rubber eraser. If necessary, clear
the water passage at the tip with a nozzle needle.
6.3.2
Manual cleaning of the interior
To effectively set up, the inside of the machine must be cleaned automatically in a
cleaning and disinfection unit in accordance with ISO 15883-1.
(The interior of this product is not to be cleaned manually).
6.3.3
Mechanical cleaning of the exterior and interior
Note
Before setting up, screw the SONICflex tips clockwise onto the SONICflex instru‐
ment.
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g. Miele G 7781 / G 7881.
(Validation was performed with the program "VARIO-TD", the cleaner "neodisher
®
mediclean", the neutraliser "neodisher
®
Z" and rinse "neodisher
®
mielclear" and
only refers to the material compatibility with KaVo products.)
Note
Insert the instrument with tip screwed on into the adapter for interior cleaning in
the thermodisinfector (e.g. Miele accessories for dental silicon adapters).
▶ The program settings and cleansers and disinfectants that must be used can
be found in the instructions for use of the thermodisinfector.
6.4
Disinfection
CAUTION
Malfunctioning from using a disinfectant bath or disinfectant containing chlorine.
Defects in the product.
▶ Only disinfect in a thermodisinfector or manually.
Instructions for use For SONICflex tips paro - REF 0.571.0371, paro A - REF 1.006.2020
6 Preparation methods according to ISO 17664 | 6.4 Disinfection