KAUKO XI-M User Manual Download Page 8

 

 

 

8 / 14 

 

 

Depending on the medical purpose of the overall system, it may also be necessary to install a 
galvanic isolator between an active medical product and computer system for the Kauko Xi-M 
computer to comply with the requirements of IEC/EN 60601-1. A galvanic isolator isolates the 
wanted signal of the respective PC interface such as network, USB or DVI interfaces. Contact 
Kauko sales or your authorized Kauko dealer for validated isolators. 

Safety instructions for operation in the patient environment 

Under national legislation, standards and guidelines are always to be followed. If the Kauko Xi-M 
computer is used in the patient environment in conjunction with active medical products that were 
introduced according 94/42/EEC guidelines, a new medical system is created and must be re-
evaluated. Peripheral devices within the patient environment, such as printers or cameras, must 
also comply with the requirements of IEC 60601-1 and IEC 60601-1-2 or be additionally equipped 
with two-stage protection for the patients. This additional protection can be ensured by using 
isolation transformers or isolators in the connecting cable. Grossly negligent non-compliance with 
current regulations places the lives of patients and those of uninvolved third parties in possible 
danger. When in doubt, contact your Kauko support team to create a safe environment for you and 
your patients. We will be happy to help you. 

3. Integration in a medical system 

In the case of integration of the Kauko Xi-M computers in a medical electrical system, the 
requirements of the valid version of IEC 60601-1 and possibly the requirements of the medical 
product guidelines 93/42 EEC (2007/47/EC) are to be complied with, and the risk evaluation must 
be done by the system creator. The Kauko Xi-M computer must only be electrically connected and 
operated with products and devices that comply with the requirements of the standards EN/IEC 
60601-1 and EN/IEC 60950-1. Before the integration of Kauko Xi-M computers in a system, be 
sure to ask the manufacturer of the medical system if this requirement for the operation with a 
computer is specified. If operation with the computer is permitted, keep strictly to the system 
manufacturer

’s recommended performance values for the computer unit. Connected data cables 

must be separately isolated, where appropriate, as well as the connecting cable between the 
Kauko Xi-M computer and other possible connected devices. 

4. Operation in a network 

The Kauko Xi-M computer may only be operated in a network that complies with the requirements 
of EN 50173 and ISO/IEC 11801. We, as the manufacturer of Kauko Xi-M computers, will not be 
liable for transmission failures, data loss or electrical surges and short circuits in your network 
structure that could occur due to non-standard installation. Only allow the installation on your 
network to be carried out by the network administrator or a person authorized by them. Do not 
carry out any changes to your network without the authorization of the network administrator. In 
addition, in the case of use in the vicinity of patients, the instructions listed under Chapter 2. 
“Patient environment” are to be taken into account. 

5. Safe operation 

Data protection 

Summary of Contents for XI-M

Page 1: ...1 14 KAUKO XI M COMPUTER User manual ...

Page 2: ...densation 5 Choice of location 6 2 Patient environment 6 Electrical safety according to IEC EN 60601 1 6 Safety instructions for operation in the patient environment 7 3 Integration in a medical system 7 4 Operation in a network 7 5 Safe operation 7 Data protection 7 Protection from power outages 8 Care and maintenance 8 Cleaning the housing 8 6 Liability and warranty 9 Contact and service 9 Discl...

Page 3: ...term stability a very efficient passive direct copper cooling system has been developed for the Kauko Xi M models Thanks to the closed aluminum housing no dust or dirt can get into the housing which means it is no longer necessary to open the housing for maintenance The computer s simple and stylish design makes it hygienic and easy to clean Its low maintenance long life cycle and high performance...

Page 4: ...4 14 ...

Page 5: ... validated by Kauko may be used If any visible damage corrosion heat sparks or smoke is detected in the mains power line the cable should be replaced immediately with a new one that has been validated by Kauko With the transfer or resale of the unit the same procedure must be followed If these instructions are not complied with the unit loses its conformity and there is a potential danger to the l...

Page 6: ...on from Kauko s support personnel and only components that have been tested and validated by Kauko may be used An unauthorized change to the unit s configuration will result in an immediate loss of conformity and can endanger the health of the user and or the patients and other people as electrical safety can no longer be guaranteed Ingress of foreign objects and liquids With any unintended intrus...

Page 7: ...has a unique VESA mounting solution integrated on the housing 2 Patient environment To protect your patients from electrical influences prevent contact between the patients and the computer housing The Kauko Xi M is a computer that was developed and tested for use in the vicinity of patients It fulfills the requirements of the EN IEC 60950 and EN IEC 60601 1 2 standards and the power supply fulfil...

Page 8: ... a medical electrical system the requirements of the valid version of IEC 60601 1 and possibly the requirements of the medical product guidelines 93 42 EEC 2007 47 EC are to be complied with and the risk evaluation must be done by the system creator The Kauko Xi M computer must only be electrically connected and operated with products and devices that comply with the requirements of the standards ...

Page 9: ... In addition to these unit related maintenance recommendations the legal requirements of national accident prevention regulations for electrical installations and equipment must be complied with Cleaning the housing You can clean the housing and power supply of your Kauko Xi M computer with a clean lint free and anti static cloth Always ensure that the computer is switched off and the mains supply...

Page 10: ...proper handling incorrect operation faulty connection non compliance with legal requirements and or external influences There is no entitlement to an indemnity or compensation for loss of use No liability for the functionality of the software can be assumed The person and or company responsible must ensure that the software used has been verified and validated We are not liable for any possible ad...

Page 11: ... supply connector 2 RJ45 Ethernet connector 3 Mini Display Port connectors 4 USB 3 connectors 5 WiFi antenna connector 6 Ground connector 7 Serial port 8 USB 2 connectors 9 Front panel USB 3 connectors 10 Front panel audio connector 3 5mm 11 Power button ...

Page 12: ...12 14 Dimensional drawing Identification Marking plate can be found on the bottom plate of the computer ...

Page 13: ...Intel AC WLAN optional Housing Anodized aluminum unibody VESA100 mount through Power supply External 12V medical PSU Graphic Integrated graphics support for processors with Intel Graphics Technology Sound system Intel High Definition Intel HD Audio via the Mini Display Port 1 2 interfaces through the processor Realtek HD Audio via a stereo microphone headphone 3 5 mm jack on the front panel LAN Gi...

Page 14: ...Kauko Oy Itämerenkatu 5 00180 Helsinki PL 116 00520 Helsinki P 358 9 5211 Fax 358 9 521 6641 info kauko com www kauko com ...