Connection of peripheral devices 

Read the operating instructions of the devices to be connected very carefully and follow the 
installation instructions of the respective products. The cable lengths of the devices must not 
exceed 3 meters, otherwise correct operation cannot be guaranteed. Make sure that connectors 
are sealed properly. 

Warning: 

With devices that provide the Kauko Xi-M with additional power via the connecting plug, 

the supply must also be tested and certified according to EN60601-1.

 

Upgrade and installation of components 

Remove the mains supply before opening the housing! 

Any opening of the housing is permitted only by one of 

Kauko’s authorized service partners and 

authorized technicians. According to applicable legal requirements, after opening the Kauko Xi-M, 
a new safety test must be carried out in accordance with EN60601-1. If you open a Kauko Xi-M 
unit, you must make a new safety test in accordance with EN60950. 

Before opening the housing, it is mandatory to remove the mains supply plug, all communication 
lines and to wait approx. 2 minutes until the electrical charge inside the unit has completely 
discharged. Otherwise, there is the potential for a serious safety hazard! Any change or 
modification to the configuration of the unit requires authorization from Kauko

’s support personnel, 

and only components that have been tested and validated by Kauko may be used. An 
unauthorized change to the unit

’s configuration will result in an immediate loss of conformity and 

can endanger the health of the user and/or the patients and other people, as electrical safety can 
no longer be guaranteed! 

Ingress of foreign objects and liquids 

With any unintended intrusion of foreign objects or liquids or in the event of condensation on the 
unit, it must immediately be separated from external power sources by removing the mains plug 
and communication lines. Under no circumstances should the unit be switched on again; instead, 
contact your authorized Kauko dealer or Kauko support team.   

Adherence to physical limit values 

In accordance with the technical data, maintain the specified permissible operating values for 
temperature, humidity and air pressure with no exceptions, as the Kauko Xi-M computer would 
otherwise run outside its specifications, and safe operation can no longer be guaranteed. If the 
computer is used outside its specified usage limits, the unit must be immediately switched off and 
the mains supply plug removed. 

Temperature fluctuations and condensation 

If the unit is exposed to large temperature fluctuations, such as those that can occur in 
transportation or storage in an unheated location, under no circumstances should the computer be 
connected straight away, as condensation moisture can accumulate internally. Wipe off visible 

Summary of Contents for XI-M

Page 1: ...1 14 KAUKO XI M COMPUTER User manual ...

Page 2: ...densation 5 Choice of location 6 2 Patient environment 6 Electrical safety according to IEC EN 60601 1 6 Safety instructions for operation in the patient environment 7 3 Integration in a medical system 7 4 Operation in a network 7 5 Safe operation 7 Data protection 7 Protection from power outages 8 Care and maintenance 8 Cleaning the housing 8 6 Liability and warranty 9 Contact and service 9 Discl...

Page 3: ...term stability a very efficient passive direct copper cooling system has been developed for the Kauko Xi M models Thanks to the closed aluminum housing no dust or dirt can get into the housing which means it is no longer necessary to open the housing for maintenance The computer s simple and stylish design makes it hygienic and easy to clean Its low maintenance long life cycle and high performance...

Page 4: ...4 14 ...

Page 5: ... validated by Kauko may be used If any visible damage corrosion heat sparks or smoke is detected in the mains power line the cable should be replaced immediately with a new one that has been validated by Kauko With the transfer or resale of the unit the same procedure must be followed If these instructions are not complied with the unit loses its conformity and there is a potential danger to the l...

Page 6: ...on from Kauko s support personnel and only components that have been tested and validated by Kauko may be used An unauthorized change to the unit s configuration will result in an immediate loss of conformity and can endanger the health of the user and or the patients and other people as electrical safety can no longer be guaranteed Ingress of foreign objects and liquids With any unintended intrus...

Page 7: ...has a unique VESA mounting solution integrated on the housing 2 Patient environment To protect your patients from electrical influences prevent contact between the patients and the computer housing The Kauko Xi M is a computer that was developed and tested for use in the vicinity of patients It fulfills the requirements of the EN IEC 60950 and EN IEC 60601 1 2 standards and the power supply fulfil...

Page 8: ... a medical electrical system the requirements of the valid version of IEC 60601 1 and possibly the requirements of the medical product guidelines 93 42 EEC 2007 47 EC are to be complied with and the risk evaluation must be done by the system creator The Kauko Xi M computer must only be electrically connected and operated with products and devices that comply with the requirements of the standards ...

Page 9: ... In addition to these unit related maintenance recommendations the legal requirements of national accident prevention regulations for electrical installations and equipment must be complied with Cleaning the housing You can clean the housing and power supply of your Kauko Xi M computer with a clean lint free and anti static cloth Always ensure that the computer is switched off and the mains supply...

Page 10: ...proper handling incorrect operation faulty connection non compliance with legal requirements and or external influences There is no entitlement to an indemnity or compensation for loss of use No liability for the functionality of the software can be assumed The person and or company responsible must ensure that the software used has been verified and validated We are not liable for any possible ad...

Page 11: ... supply connector 2 RJ45 Ethernet connector 3 Mini Display Port connectors 4 USB 3 connectors 5 WiFi antenna connector 6 Ground connector 7 Serial port 8 USB 2 connectors 9 Front panel USB 3 connectors 10 Front panel audio connector 3 5mm 11 Power button ...

Page 12: ...12 14 Dimensional drawing Identification Marking plate can be found on the bottom plate of the computer ...

Page 13: ...Intel AC WLAN optional Housing Anodized aluminum unibody VESA100 mount through Power supply External 12V medical PSU Graphic Integrated graphics support for processors with Intel Graphics Technology Sound system Intel High Definition Intel HD Audio via the Mini Display Port 1 2 interfaces through the processor Realtek HD Audio via a stereo microphone headphone 3 5 mm jack on the front panel LAN Gi...

Page 14: ...Kauko Oy Itämerenkatu 5 00180 Helsinki PL 116 00520 Helsinki P 358 9 5211 Fax 358 9 521 6641 info kauko com www kauko com ...

Reviews: