28
Let us now understand the meaning of each symbol.
We are now familiar with the graph
zone, which shows the outline of the
crucible temperature as a function of
time.
The fusion step being edited is
represented as the portion of curve
between the two vertical red lines.
The
large
crucible
with
red
thermometer represents a heating
step, and the numeral “4” indicates
that we are now viewing the
parameters of step “Heating 4”.
The Left and Right arrows are used to
scroll among the steps of the current
program.
The
small
crucible
with
red
thermometer icon is placed just
besides a cell where the target crucible
temperature is displayed. (The Minus
and Plus buttons will be used to
change the parameter.)
The stopwatch icon is placed just
besides a cell where the step duration
is displayed. Note that step durations
are calculated
after
the required
temperature is reached.
The stopwatch with arrow icon is
placed just besides a cell that indicates
how fast the target temperature will be
reached.
Most fusion applications can use the
“Fast” setting, but oxidation steps
often call for a slower heat-up rate.
This is called “ramping”, and its speed
is expressed in °C/minute.
Summary of Contents for K1 PRIME
Page 1: ...Automatic fluxer 1...
Page 2: ......