44
Cables which have been shown to comply with IEC 60601-1-2:
Cable type
Shielded
Length [m]
Ferrite
Used for
PA
No
> 3
No
Potential equalization
SCB Cable
Yes
>
1
No
Connecting SCB network, other SCB units
20
0141 30
Yes
3
No
One-pedal-Footswitch
27
6100 70
Yes
0.42
No
Control cable (connecting cable between the S-PILOT™ and
the UNIDRIVE
®
S III)
20
7010 70
Yes
1
No
Control cable (connecting cable between the S-PILOT™ and
UNIDRIVE
®
S III /CALCUSON)
UP 004
No
3
No
Connecting cable between the S-PILOT™ and
AUTOCON
®
III 400/ConMed system 2450 or 5000
UP 005
No
3
No
Connecting cable between the S-PILOT™ and Covidien
Force Triad or Covidien Valleylab Force FX
Power Cord
No
3
No
Power supply
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The S-PILOT™ model UP 501 is intended for use in the electromagnetic environment speci
Ƃ
ed below.
The customer or user of the
S-PILOT™
model UP 501 should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The S-PILOT™ model UP 501 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The S-PILOT™ model UP 501 is suitable for use in all establishments including
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Electromagnetic
Compatibility (EMC)
information