Instructions for use SOM 62/32/22 LED
SOM 62/32/22 LED EN 2017-03
10
7.2
Installation of SOM 32
7.2.1
Installation of the ceiling bracket
Carry out the following installation steps as in Figure 3:
Using the aligner (1), drill four holes into the ceiling (Ø 10mm, hole depth 80mm)
Clean the holes and check the distance between the holes using the aligner (1)
Hammer in four through anchors (2) up to the mark and secure them to the torque specified
(45 Nm)
Secure the aligner (1) to the ceiling using four screws (3) and four nuts (4)
Before fitting the ceiling column to the aligner, place the mains cable (15) (on the connector
side) into the groove and ensure the mains cable is not pinched
Unscrew the socket plate (6) from the column. Pull the cable down by about 150mm.
Screw four threaded bolts (7) into the aligner (1) and secure them with lock ring (8) and nut
(9). Screw another four nuts (9) onto the threaded bolts (7). Put the washers (11) onto the
threaded bolts.
Insert the column with flange (10) into the four threaded bolts (7) of the aligner, and secure
with four lock rings (8), four washers (11) and four nuts (9). Then align the column to be
perfectly vertical.
Put on the cover (12), guide the mains plug out of the ceiling bracket and attach it with ring
(13) and three threaded pins (14)
Carefully guide the cable (15) into the column, and insert and tighten the socket plate (6)
Figure 3