The CPR Valve on the mattress can also be used for a CPR procedure by rotating the CPR Valve to the
open position
•
To deflate the FOAM AIR mattress for CPR procedure, disconnect the mattress deflate connector which
is attached to the “DEFLATE” tag from the deflate valve. If connected to a control unit, press the quick
release buttons on the connector coupling bodies and simultaneously pull the hose from the control unit
flange connector.
PATIENT TRANSPORTATION
•
To transport a patient without removing the patient off the bed, turn off the control unit, disconnect the
power cord from the power source and roll it up on the control unit securely.
SKINGUARD® FLOAT CONVERTIBLE FOAM AIR MATTRESS AS A NON-POWERED MATTRESS
SKINGUARD® Float Convertible mattress can be used as a non-powered mattress without the control unit.
Before using the mattress connect the control unit to the mattress and inflate the mattress by setting the control
unit comfort level to 5. Once the control unit fills the mattress and the pressure stabilizes then turn the unit off
and disconnect the hose assembly from the mattress and control unit and store the control unit and the hose
assembly in a storage area.
SKINGUARD® FLOAT CONVERTIBLE FOAM AIR MATTRESS AS A POWERED MATTRESS
If need arises the
SKINGUARD® FLOAT CONVERTIBLE foam air mattress can be converted into a powered
mattress by simply connecting the control unit to the mattress using the hose assembly provided with the
system. For operating instructions in powered mode please refer to
OPERATING INSTRUCTIONS ABOVE
FOR SKINGUARD® FLOAT
SKINGUARD® TURN
POWER
•
During initial power up (when power cord is plugged into the power source), the control unit will go
through a system initialization routine for a few seconds and enter “STANDBY” (amber LED on) mode.
•
If the unit is in STANDBY mode with amber LED on, press the POWER button to turn the control unit on.
•
If the power comes on after a power outage, the control unit will go through its system initialization routine
for few seconds and then resume the desired function.
MAX INFLATE
•
Press the MAX INFLATE button and the unit will rapidly inflate the mattress at maximum flow.
•
The mattress will inflate to its normal size within 60 seconds (inflation time depends on the size of the