ALTERNATING PRESSURE   

 

  

  To set Alternating Pressure Mode, press the AP TIMES button and select the desired time for the 

alternating pressure therapy; LED lights up. The AP times are 5, 10, 15, and 20 minutes. 

 

  In the Alternating Pressure Mode, alternate air cushions will inflate to the patient comfort pressure while 

those cushions between the inflated air cushions will deflate 0% or 50% (factory set) of the set patient 
comfort pressure. An AP cycle will alternate these inflated and deflated air cushions, switching back and 
forth. 

 

FOWLER    

    

  To activate Fowler Mode, press the Mode button until the LED light is turned on next to Fowler 

 

  In this mode, pressures in the entire mattress will be increased to higher than the set comfort pressure 

level (not exceeding 32mmHg). This enables the patient in a fowler position to be supported without 
bottoming out. 

 

  When the Auto Fowler is activated, the LED will light up. 

 

LOW AIR LOSS   

 

 

  For the SkinGuard Float®

press the LAL / MAX INFLATE until the LED is turned on near the LAL 

description to active. 

 

LOCK OUT   

   

 

  Control unit functions (including power) can be completely locked, preventing tampering of settings by 

pressing and holding the LOCK button until the light comes on (approximately 3~5 seconds). Press and 
hold the LOCK button to unlock. 

 

ALARM SILENCE    

   

 

  An audio-visual alarm is sounded in the event of power failure or when the hose is disconnected from the 

unit. Audio alarm can be muted by pressing alarm silence key. 

 
 
 

Summary of Contents for SKINGUARD

Page 1: ...SKINGUARD MATTRESS SERIES OWNER S MANUAL Please read this manual before using this product Do not discard Save for future reference This manual must be given to the user of this product ...

Page 2: ... 340 4361 Customer Service 1 866 KAP MED 1 866 527 6331 Technical Support Service 1 951 340 4360 Email sales kapmedical com service kapmedical com www kapmedical com FDA Registered Company California FDA Registered Company Medicare Coded SADMERC Health Canada Medical Device Licensed ISO 13485 Certified Company Medical Device Electrical Safety Tested SGS Fire Safety Tested MADE IN USA ...

Page 3: ...anation of Symbols 10 Product Overview 12 Control Unit Features 12 Support Surface Features 13 SkinGuard Quick Reference Guide 15 SkinGuard Float Quick Reference Guide 16 SkinGuard Turn Quick Reference Guide 17 Unpacking the System 18 System Setup 19 Operating Instructions 20 Cleaning Procedure 32 Care and Storage 34 Troubleshooting Guide 35 Preventive Maintenance 36 Accessories 36 Warranty Inform...

Page 4: ...g to qualified service personnel Equipment should only be connected to a properly grounded three pronged wall outlet using the 14 foot 427 cm hospital grade power cord provided with the product Warning CONTRAINDICATIONS ALWAYS consult the patient s physician before using the SkinGuard Mattress System Never drop or insert objects into any opening of the control unit Please read this manual before o...

Page 5: ...rol unit Care should be taken such that the control unit is not blocked and kept away from any heat source or radiators during the operation of the unit Care should be taken such that the power cord of the control unit is not pinched or has any objects placed on it Make certain it is not located where it can be stepped on or tripped over Do not attempt to service the control unit except as explain...

Page 6: ... rail the rail and the side edge of the headboard or foot board RISK OF ENTRAPMENT Bed Rail Entrapment is a known risk in the use of bed s equipped with bed rails Every patient is unique Only the patient s medical care provider is familiar with the patient s unique medical condition and needs Only the patient s medical care provider and or the dealer from whom you obtained this equipment upon prop...

Page 7: ...l increased risk of entrapment The risk judgment must be performed by a medical professional The U S Food and Drug Administration in partnership with the U S Department of Veterans Affairs Health Canada s Medical Devices Bureau and representatives from national health care organizations and provider groups patient advocacy groups and medical bed and equipment manufacturers including the Hospital B...

Page 8: ...D 1 R Breaking Capacity BRK CAP SkinGuard SkinGuard Turn BRK CAP 125 VAC is 10kA 250 VAC is 200A SkinGuard Float BRK CAP 125 VAC is 10kA 250 VAC is 35A Mode Of Operation Continuous PERFORMANCE SPECIFICATIONS Weight Capacity Standard Mattress 500 Lb 226 Kg maximum Bariatric Mattress 1000 Lb 455 Kg maximum Mattress Overlay 250 Lb 113 kg maximum U S INTL Pressure Zones 5 AP Zones 2 Rotation Zones 2 M...

Page 9: ...nd washable Cal 117 pass Base Non skid embossed PVC nylon knit Optional CFR 1632 CFR 1633 compliant mattresses are available Top Sheet DERMA PLUSH urethane transfer coated polyester knit low friction low shear force producing breathable liquid resistant and highly vapor permeable Cal 117 pass Description Inflated Dim LxWxH Weight Overlay 80 x36 x3 or 5 8 lbs 203x89x13cm 3 6 Kg Standard Mattress 80...

Page 10: ...ard for safety of Medical Electrical Equipment Flame Resistance Unit components meet UL 94V 0 Mattress components pass California117 Optional California TB 106 TB 129 16 CFR 1632 16 CFR 1633 BS 6607 CRIB 5 BS 597 1 BS 597 2 compliant mattresses lined with Kevlar fire barrier available Kevlar lining based on flammability standard FDA REGISTRATION FDA registered company as a manufacturer and as a co...

Page 11: ...patient comfort pressure level AP TIMES Selects Alternating Pressure therapy and sets Alternating Pressure times 5 10 15 20 minutes PULSE The pressure in each air cell will be increased by 40 for 30 seconds and then reduced to the prior pressure setting for 30 seconds This cycle will continue every minute as long as pulsation is enabled MAX INFLATE Inflates mattress rapidly 15 minute timer LOW AIR...

Page 12: ...will sound Indicates the point of attachment of the equipment to earth grounding point Attention Instructs end user care giver operator to refer to the manual Indicates that the degree of protection against electrical shock is TYPE BF Not for use in presence of flammable anesthetics Consult Instructions for Use Waste electrical and electronic equipment recycle Risk of electrical shock Do not remov...

Page 13: ...y system is comprised of a durable zippered PVC nylon knit base and top sheet which houses a urethane coated nylon durable 5 inflated air pad in the form of 10 to 20 fixed air cushions with hose assembly The mattress replacement system is comprised of a durable PVC nylon knit base with a safety 2 convoluted air base 5 or 8 inflated detachable air cushions and covered with a vapor permeable water p...

Page 14: ...h dual fuse not shown in picture Power Fail LED flashes to indicate power outage Low Pressure L P flashes on display panel to indicate low pressure Lock Switch to lock out all control functions Support Surface Mattress Overlay not all features available with each model SKINGUARD SKINGUARD FLOAT SKINGUARD TURN Self contained mattress replacement system mattress overlay system with easily detachable...

Page 15: ... convenience of protecting the power cable from accidental damage The mattress includes a CPR Valve at the bottom of the mattress for rapid deflation Features SkingGuard SkinGuard Float SkinGuard Turn Low Air Loss Alternating Pressure Pulsation Heel Zone Regulator Continuous Lateral Rotation Mattress Options SkingGuard SkinGuard Float SkinGuard Turn Standard 80 L x 36 W x 8 H Optional 10 height Ex...

Page 16: ...by 40 not exceeding 32mmHg for 30 seconds and then reduced to the prior pressure setting for 30 seconds This cycle will continue every minute as long as pulsation is enabled MAX INFLATE Press the MAX INFLATE button LED on and the mattress will rapidly inflate Beeps will occur every 3 minutes as a reminder Max Inflate is active Max Inflate will deactivate after 15 minutes FOWLER Press the FOWLER bu...

Page 17: ...osition In A P mode alternate air cushions will inflate to the patient comfort pressure while those cushions between the inflated air cushions will deflate 0 or 50 factory set of the set patient comfort pressure LOW AIR LOSS MAX INFLATE Press the LAL MAX INFLATE until the LED is turned on near the desired mode description to active that mode In MAX INFLATE mode the mattress will rapidly inflate Be...

Page 18: ... will deactivate after 15 minutes SELECT PRESS the SELECT button to choose between Therapy Pulse Multi and Fowler Upright modes TURN Press the TURN button to select Turn Mode Follow the instructions on the right side of the screen to adjust the Left Turn Right Turn Dwell time Left Degree and Right Degree of turn settings PAUSE Press the PAUSE button to pause the turn cycle therapy time Press PAUSE...

Page 19: ...d System Box 1 Control Unit Box 1 Mattress Control Unit Box 1 Control Unit 1 Operating Instruction Owner s Manual 1 Power Cord Unpacking and Inspection Before accepting and signing for your shipment please inspect the box or boxes for external and internal damages Verify that the number of boxes listed on the packing list matches the number of boxes received Verify that no components are damaged o...

Page 20: ...em There are ten nylon black straps with buckles two straps at the head of the mattress two on the foot of the mattress and three on the each side of the mattress shown Loop each strap around the bed frame and fasten it securely to the bed frame using the buckle 6 Extend the hanger on the back of the control unit and suspend the control unit from the footboard of the bed If the bed you are using d...

Page 21: ...e the desired function MAX INFLATE Press the MAX INFLATE button and the unit will rapidly inflate the mattress at maximum flow The mattress will inflate to its normal size within 60 seconds inflation time depends on the size of the mattress In this mode the entire mattress will be pressurized to 35 5 mmHg During this mode a series of beeps will sound every 3 minutes as a reminder that Max Inflate ...

Page 22: ...inflated and deflated air cushions switching back and forth PULSE To set Pulse Mode press the PULSE button LED lights up For SkinGuard Turn press the SELECT button until the Pulse mode is displayed on the screen The pressure in each air cell will be increased by 40 for 30 seconds and then reduced to the prior pressure setting for 30 seconds This cycle will continue every minute as long as pulsatio...

Page 23: ... signal to alert the caregiver by flashing the amber POWER FAIL LED and the buzzer will turn on Once the power is restored to the control unit the audiovisual signal will cease and the unit will resume operating its set mode During power outage the mattress will retain air as long as the mattress is connected to the control unit HEEL ZONE REGULATOR Heel Zone Pressure can be adjusted independently ...

Page 24: ...The width expansion regulator valve is located at the patient s right foot end of the mattress and the length expansion regulator valve is located at the patient s left foot end of the mattress These valves allow the mattress to expand in width and length based on selected setting If the Expandable Mattress System is an Expandable Length System only the length expansion valve can be found on the m...

Page 25: ...e screen flashing the amber POWER FAIL LED and turning on the buzzer Once the low pressure problem is stable the audiovisual signal will cease and the unit will resume operating its set mode CPR FUNCTION To deflate the mattress overlay pad or for a CPR procedure disconnect the mattress hose from the control unit by pulling the mattress hose connector straight out from the control unit magnetic con...

Page 26: ... the desired mode of MAX The mattress will inflate to its normal size within 15 minutes inflation time depends on the size of the mattress In this mode the entire mattress will be pressurized to 35 5 mmHg During this mode a series of beeps will sound every 3 minutes as a reminder that Max Inflate Mode has been activated PATIENT COMFORT CONTROL LEVEL Pressing the buttons will decrease or increase t...

Page 27: ...LED light is turned on next to Fowler In this mode pressures in the entire mattress will be increased to higher than the set comfort pressure level not exceeding 32mmHg This enables the patient in a fowler position to be supported without bottoming out When the Auto Fowler is activated the LED will light up LOW AIR LOSS For the SkinGuard Float press the LAL MAX INFLATE until the LED is turned on n...

Page 28: ...hing the amber POWER FAIL LED and turning on the buzzer Once the low pressure problem is stable the audiovisual signal will cease and the unit will resume operating its set mode EXPANDABLE MATTRESS SYSTEM XMS EXPANDABLE LENGTH MATTRESS SYSTEM EL IF EQUIPPED Mattress has ability to expand in width from 36 42 48 and transport mode and or in length 80 84 and 88 by using the mattress expansion regulat...

Page 29: ...EL make sure to tighten or loosen the straps accordingly to the mattress width EXPANSION AIR CELLS To increase the length of the mattress obtain an air cell expansion kit and connect the male connector to the female connector on the exterior wall at the end of the mattress see picture above Use the straps and snaps to secure the expansion top sheet to the mattress CPR FUNCTION To deflate the mattr...

Page 30: ...urn the unit off and disconnect the hose assembly from the mattress and control unit and store the control unit and the hose assembly in a storage area SKINGUARD FLOAT CONVERTIBLE FOAM AIR MATTRESS AS A POWERED MATTRESS If need arises the SKINGUARD FLOAT CONVERTIBLE foam air mattress can be converted into a powered mattress by simply connecting the control unit to the mattress using the hose assem...

Page 31: ...ns on the screen in each menu to select desired settings PATIENT SETUP SkinGuard Turn only To set the patient s height and weight press the PATIENT SETUP button In this menu use the ADJUST SETTINGS arrows to enter the patient s weight and height Use the SELECT button to accept the weight and height values Once set the mattress will inflate to the recommended comfort pressure level This is a recomm...

Page 32: ...he microprocessor will activate an audiovisual signal to alert the caregiver by flashing the amber POWER FAIL LED and the buzzer will turn on Once the power is restored to the control unit the audiovisual signal will cease and the unit will resume operating its set mode During power outage the mattress will retain air as long as the mattress is connected to the control unit LOW PRESSURE L P flashi...

Page 33: ... of mattress allow the UM mattress to be inflated Follow instructions provided on the flap covering the valve PATIENT TRANSPORTATION ON AN UMS MATTRESS To transport a patient without removing the patient off the bed turn off the control unit disconnect the power cord from the power source and roll it up on the control unit securely The mattress will deflate within a few seconds after the control u...

Page 34: ...he cleaning operations are completed remove and dispose the rubber gloves appropriately Wash your hands thoroughly with antibacterial soap MATTRESS The complete support surface of the mattress is made of superior quality materials and is modular in construction All the components such as manifold hose assembly air cushions top sheet air pad cover and mattress base are interchangeable and can be ea...

Page 35: ...can be cleaned following the steps below 1 Wear eye goggles and protective gloves before starting the cleaning procedure 2 Follow steps 2 through 4 in control unit cleaning procedure above to prepare disinfectant solution 3 Clean the top and bottom mattress cover using the prepared disinfectant solution and refer to step 4 of the above Air Mattress cleaning instructions for washing instructions 4 ...

Page 36: ... or replace split hose 4 Replace leaking air cushions or overlay pad 5 Un kink manifold or replace split Manifold 6 Send control unit back to factory for repair 7 Send control unit back to factory for repair 8 Send control unit back to factory for repair B No Power 1 Control Unit OFF 2 Power cord disconnected 3 No power in the power source 4 Power outage 5 Blown fuse 1 Check power source and turn ...

Page 37: ...rformed only by factory authorized and qualified technical personnel FILTER CLEAN FILTER EVERY 3 MONTHS OR WHENEVER DIRTY Remove the 2 thumb screws from the filter grill and separate filter foam Wash filter foam using soap and water wash and replace filter back on the unit and fasten the thumb screws ACCESSORIES SkinGuard Control Unit only SkinGuard Mattress only SkinGuard Float Control Unit only ...

Page 38: ...er no circumstances shall KAP Medical be liable for any loss direct indirect incidental or special damages arising out of or in connection with the use of this product The control unit warranty does not cover normal maintenance such as cleaning periodic electrical tests performance tests and updating of equipment or parts thereof This warranty shall be void and not apply if the control unit includ...

Page 39: ...intenance and Repair Log UNIT SERIAL _____________________________________ MATTRESS SERIAL _______________________________ AUTO FOWLER SERIAL ____________________________ DATE PURCHASED _________________________________ Date Air Filter Power Cord Mattress Repair C cleaned OK okay R repaired replaced specify ...

Reviews: