SkinGuard® Series
Owner’s Manual
PN 400901 Rev. C
ECO 18-0281
Page 27
POWER FAIL
•
In the event of a power outage, the microprocessor will activate an audiovisual signal to alert the
caregiver by flashing the amber “POWER FAIL” LED and the buzzer will turn on. Once the power is
restored to the control unit, the audiovisual signal will cease and the unit will resume operating its set
mode. During power outage the mattress will retain air as long as the mattress is connected to the control
unit.
LOW PRESSURE (
“L” “P” flashing on display panel)
•
In the event of hose disconnection, the microprocessor will activate an audiovisual signal to alert the
caregiver either by flashing “L” “P” on the display panel and turning on the buzzer or displaying Low
Pressure on the screen, flashing the amber “POWER FAIL” LED, and turning on the buzzer. Once the
low pressure problem is stable, the audiovisual signal will cease and the unit will resume operating its set
mode
EXPANDABLE MATTRESS SYSTEM (XMS) / EXPANDABLE LENGTH MATTRESS SYSTEM (EL)
(IF EQUIPPED)
Mattress has ability to expand in width from 36” (42”, 48”, and transport mode) and/or in length (80”, 84” and 88”)
by using the mattress expansion regulator valves.
The Expandable Mattress System (XMS) has two (2) expansion valves; one located in the right and the other in
the left side, foot end of the mattress. The width expansion regulator valve is located at the patient’s right, foot
end of the mattress and the length expansion regulator valve is located at the patient’s left, foot end of the
mattress. These valves allow the mattress to expand in width and length based on selected setting. If the
Expandable Mattress System is an Expandable Length System, only the length expansion valve can be found on
the mattress.