13. Troubleshooting
13.1 Batteries
Symptom
Possible Cause(s) / Solution(s)
Refer To
(Page No.)
Nothing appears on the screen
when the power is turned ON.
The batteries may be low.
→
Turn the power OFF and replace the batteries.
6
Nothing appears on the screen
after new batteries are installed.
Batteries may be installed with incorrect polarity.
→
Turn the power OFF and install the batteries correctly.
6
“E--0” is displayed on the
screen.
The batteries may be low.
→
Turn the power OFF and replace the batteries.
6
13.2 Initial Operation
Symptom
Possible Cause(s) / Solution(s)
Refer To
(Page No.)
“E--9” is displayed on the
screen.
ROM Cassette is not installed.
→
If the ROM Cassette is already installed, turn the power OFF
and inspect to ensure ROM is properly seated.
8
“----” is displayed on the screen. Probe is not connected properly.
→
Turn the power OFF, and connect the probe properly.
7
The measurement unit is not
correct.
You can change the Air Velocity unit (m/s
⇔
FPM) and the Air
Temperature unit (
℃⇔
o
F).
29
13.3 During Operation
Symptom
Possible Cause(s) / Solution(s)
Refer To
(Page No.)
Reading is displayed as “----”.
Measurable range is exceeded.
→
Use the instrument within the specified measurement range.
32
Probe/Probe Cable may not be connected property.
→
Check probe connection.
7
Probe/Probe Cable wire is disconnected or the sensor is damaged.
→
Contact your local KANOMAX Office or service center.
Incorrect air velocity reading
Confirm that the direction mark of the probe is facing against the
airflow.
9
Air temperature reading is high. Correct reading cannot be obtained when there is no airflow.
Minimum 0.1m/s velocity is required for measurement.
9
The response for Air Velocity
Measurement is slow.
Confirm the Time Constant settings.
13
Display shows “E - - 8” when
adjusting pressure zero point.
Pressure Port (
+
,
-
) may be blocked.
→
Unblock both (
+
) and (
-
) ports.
9
The adjustment exceeds the range specified for the Zero
Adjustment.
→
Contact your local KANOMAX Office or service center.
9
13. Troubleshooting
37