4.6.1 HEATING PAD ALARM 

This alarm will be activated if the pad is not connected, or not properly connected, or if any of the  
8 (4) heating elements deviates from normal electrical resistance, or if there is a fault in the connecting 
cable, in the pad connector or in the regulator power switch. 
In case of a pad failure being detected during the self-test, the Operatherm will not warm until the pad 
failure has been corrected. 
If, during use, a pad fault occurs that affects 3 or more of the 8 heating elements the unit will alarm 
visually and acoustically and switch off the heating. On short pads the heating will be switched off if 2 or 
more of the 4 elements are faulty. 
In case the fault affects maximum 2 of the 8 elements or 1 of the short Pad elements, the unit will alarm 
visually as follows: 

 

1 element faulty - slow flashing Pad fault LED

 

 

2 elements faulty - slow double flashing Pad fault LED

 

 
If the Heating Pad gets disconnected during normal use, the pad fault indicator and the 32oC  LED will 
be lit.  The beeper will also sound. Switch the unit off and reconnect the Heating Pad. Restart the unit 
and observe a correct self test. 

 

 

4.7 CONTROL 

UNIT 

FAULT 

Any fault in the Control Unit will lit the Control Unit fault symbol and activate the acoustic alarm. The 
heating will be switched off. 
Switch the unit off and try a restart. Observe that a correct self test takes place and that the unit 
behaves normally and indicates temperatures as described in this manual. If in the slightest doubt about 
the correct function of the unit, do not use it. Contact a qualified technician. 

 

4.8 MAINS 

FAILURE 

ALARM 

In case of a mains failure, the Operatherm alarms by flashing the C. U. LED and sounding the beeper 
intermittently. 
The mains failure alarm is kept active for at least 10 min. 
After return of the mains power the Operatherm restarts with a self test and at the temperature selected 
before the mains failure occurred

 

4.9 REAR-PANEL 

The rear-panel contains information about manufacturer, serial number, voltage, fuses, etc. Mains cord 
is also connected here. 

        

FIG. 2 REAR PANEL

 

 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for Operatherm 200

Page 1: ...ENANCE 12 NOTE Patient warming equipment can harm the patient if not used correctly Never use a unit that do not perform the Self Test according to this manual Make sure that the performance of the unit is regularly checked This manual contains important information and must be kept for future reference It is valid for all Operatherm 202 200 units Manufactured by KanMed AB Gårdsfogdevägen 18B SE 1...

Page 2: ...it has been examined by a qualified technician The Operatherm 202 is a very safe equipment However certain combinations of heat pressure long term surgery and disinfectant fluids can cause skin damage Therefore it is important to observe the advice regarding pressure relief excessive fluid and temperature selections given later in this section 2 2 EXPLANATION OF SYMBOLS Equipment not suitable for ...

Page 3: ... safety 4 The Heating Pad cable may not be altered in length or repaired in any other way 5 See Users Manual section 10 concerning cleaning and maintenance of Heating Pads and Control Units 6 Sharp objects must not penetrate the Heating Pad 7 Place the Heating Pad flat on the operating table with the text side towards the patient 8 Gel Pads must be placed over the Heating Pad 9 When used with chil...

Page 4: ...ent 2 8 TEMPERATURE SETTING Under normal conditions and with KanMed Gel Pads under the patient the maximum temperature 39 C should be selected During long interventions one must consider a lower temperature setting If the OR table surface is hard blood pressure is low i e bad circulation old patients patients with diabetes etc 37 C should be the highest temperature 2 9 USE WITH CHILDREN AND PRE TE...

Page 5: ... Watertight except connector Surface electrical properties Not electrical conductive Control Unit Voltage 220 240 V AC 100 and110 120 V AC is also available Frequency 50 60 Hz Power 285 VA max Temperature range 32 C and 35 39 C in steps of 0 5 C Accuracy 0 5 C Indicators LED Resolution 0 5 C Over temperature alarm level 40 25 C 0 75 C Dimensions L 255 mm approx W 150 mm approx H 120 mm approx Weig...

Page 6: ... is switched on it will perform a self test of all functions including the internal safety circuit This is how the self test runs When pressing the ON OFF button all front panel indicators and the beeper are activated After about 0 5 sec the red ALARM indicators and the beeper are turned off About 2 sec later temperature indicators are turned off Simultaneously the red indicators and the beeper ar...

Page 7: ...long pad or of the 4 elements in the short pad should become warmer than 40 25 C The visual and acoustic alarm will be activated and the over heated element s will be switched off for 90 sec The acoustic alarm can be reset but the over temperature indicator will stay lit If during the 90 sec period the temperature of the alarming element s drops below 40 25 C the visual alarm will automatically ca...

Page 8: ...ets disconnected during normal use the pad fault indicator and the 32oC LED will be lit The beeper will also sound Switch the unit off and reconnect the Heating Pad Restart the unit and observe a correct self test 4 7 CONTROL UNIT FAULT Any fault in the Control Unit will lit the Control Unit fault symbol and activate the acoustic alarm The heating will be switched off Switch the unit off and try a...

Page 9: ...y a technician 5 3 SELECTING TEMPERATURE In case you wish to select another temperature than 37 C then do as described in chapter 4 4 1 It will take about 8 15 minutes for the elements in the pad to reach the set temperature depending on surrounding conditions NOTE Please refer to section 2 regarding choice of temperature 5 4 GEL PADS Always use Gel Pads to get an even heat distribution a greater ...

Page 10: ... 1 2 MAINS POWER FAILURE ALARM AND RESTARTING Connect the Heating Pad to the Control Unit and the Control Unit to the mains Switch on the unit and wait for the self test to be completed Disconnect the power cord The alarm shall sound for at least 10 minutes Reconnect the power cord The unit shall perform the self test and then warm at the temperature that was selected before power off 6 1 3 FAULTS...

Page 11: ...ntact your local dealer 8 EMC COMPATIBILITY STATEMENT Please see www kanmed se section downloads Operatherm 9 WARRANTY STATEMENT KanMed warrants the purchaser that the Operatherm Control Unit and Heating Pads shall be free from defects in material and workmanship for a period of 12 month from the date of purchase The sole obligation of KanMed with respect to any such defect is limited to the repai...

Page 12: ...ried out by the user operator Self test at Power ON Every time the equipment is switched on Visual inspection of pad cable and connector Before every use Checks carried out by technical personnel Alarm functions Once a year Mains failure Pad failure Over temperature Temperature control Once a year Electrical safety Once a year Exchange of connector saver Once a year 10 3 STORING The Heating Pad sh...

Reviews: