background image

Manual KanMed Operatherm 

 

Page 

5

 

 

 

 

3 TECHNICAL 

SPECIFICATIONS 

 

Heating Pad Standard  

Art. no.:OP-200-002

 

 

Voltage 

 

30 Volt DC 

 Power 

  

200 

Watt 

 

Dimension 

 

L 1040 mm  W 450mm  H 2 mm 

 Weight 

  

kg 

 

Length of cable 

3,5 meter approx. ( Other lengths may be ordered ) 

 

Water protection 

Watertight, except connector 

 

Surface, electrical properties 

Not electrical conductive 

 

Heating Pad Small 

       

 

 

Art. no.: OP-200-012 

 

Voltage              

 

 

30 Volt DC 

           Power                

 

 

100 Watt 

      

Dimension         

 

 

L 630 mm  W 430 mm  2 mm 

 Weight 

    1,5 

kg 

 

Length of cable        

 

3.5 metre approx. (Other lengths may be ordered) 

 

Water protection 

Watertight, except connector 

 

Surface, electrical properties.  

Not electrical conductive 

 

Heating Pad 150cm  

Art. no.: OP-200-221

 

 

Voltage 

 

30 Volt DC 

 Power 

  

200 

Watt 

 

Dimension 

 

L 1500 mm  W 450 mm  H 2 mm 

 Weight 

 2,8 

kg 

 

Length of cable 

3,5 metre approx 

 

Water protection 

Watertight, except connector 

 

Surface, electrical properties 

Not electrical conductive 

 

Control Unit 

  

 

Voltage              

 

 

220 -240 V AC ( 100 and110-120 V AC is also available ) 

    

Frequency            

 

 

50/60 Hz 

   

Power                

 

 

285 VA (max) 

    

Temperature range    

 

32°C and 35-39°C in steps of 0.5°C 

 

 Accuracy 

   0.5°C 

 

 Indicators 

   LED 

 

 Resolution 

   0.5°C 

 Over-temperature 

alarm 

level 40,25°C 

±

  0.75°C 

 

Dimensions       

 

 

L  255 mm approx. W  150 mm approx H  120 mm approx 

 

Weight               

 

 

5 Kg approx. 

     

Fuses                

 

 

220 - 240 Volts  T 4A,  110 –120 Volts  T 8A 

 Mode 

of 

use 

   Continuous 

 

Heating up time

 

 

 

Elements inside the Heating Pad 

 

about 8-10 minutes at room temperature 

 

Heating Pad surface

   

 

 about 25 minutes when tested under conditions of  
ADEQUATE HEAT DISCHARGE 

(IEC601-2-35)

 :   

 

Environment  

 

 

Operating temperature 

 

15°C - 40°C 

 

Shipping and storage  

 

-20°C - + 60°C 

 

Relative humidity 

 

 

10 - 85 % non condensing 

 

 

 

 

 
Safety Standards 

 

   EN60601-1, 

EN60601-1-2, 

IEC60601-2-35 

 

Protection type  

 

 

BF, Defibrillator safe 

 

Protection class  

 

 

Class I

  

 

MDD classification   

 

Class IIB 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0413

 

 

 
NOTE: Specifications are subject to changes at any time. 

 

Summary of Contents for Operatherm 200

Page 1: ...ENANCE 12 NOTE Patient warming equipment can harm the patient if not used correctly Never use a unit that do not perform the Self Test according to this manual Make sure that the performance of the unit is regularly checked This manual contains important information and must be kept for future reference It is valid for all Operatherm 202 200 units Manufactured by KanMed AB Gårdsfogdevägen 18B SE 1...

Page 2: ...it has been examined by a qualified technician The Operatherm 202 is a very safe equipment However certain combinations of heat pressure long term surgery and disinfectant fluids can cause skin damage Therefore it is important to observe the advice regarding pressure relief excessive fluid and temperature selections given later in this section 2 2 EXPLANATION OF SYMBOLS Equipment not suitable for ...

Page 3: ... safety 4 The Heating Pad cable may not be altered in length or repaired in any other way 5 See Users Manual section 10 concerning cleaning and maintenance of Heating Pads and Control Units 6 Sharp objects must not penetrate the Heating Pad 7 Place the Heating Pad flat on the operating table with the text side towards the patient 8 Gel Pads must be placed over the Heating Pad 9 When used with chil...

Page 4: ...ent 2 8 TEMPERATURE SETTING Under normal conditions and with KanMed Gel Pads under the patient the maximum temperature 39 C should be selected During long interventions one must consider a lower temperature setting If the OR table surface is hard blood pressure is low i e bad circulation old patients patients with diabetes etc 37 C should be the highest temperature 2 9 USE WITH CHILDREN AND PRE TE...

Page 5: ... Watertight except connector Surface electrical properties Not electrical conductive Control Unit Voltage 220 240 V AC 100 and110 120 V AC is also available Frequency 50 60 Hz Power 285 VA max Temperature range 32 C and 35 39 C in steps of 0 5 C Accuracy 0 5 C Indicators LED Resolution 0 5 C Over temperature alarm level 40 25 C 0 75 C Dimensions L 255 mm approx W 150 mm approx H 120 mm approx Weig...

Page 6: ... is switched on it will perform a self test of all functions including the internal safety circuit This is how the self test runs When pressing the ON OFF button all front panel indicators and the beeper are activated After about 0 5 sec the red ALARM indicators and the beeper are turned off About 2 sec later temperature indicators are turned off Simultaneously the red indicators and the beeper ar...

Page 7: ...long pad or of the 4 elements in the short pad should become warmer than 40 25 C The visual and acoustic alarm will be activated and the over heated element s will be switched off for 90 sec The acoustic alarm can be reset but the over temperature indicator will stay lit If during the 90 sec period the temperature of the alarming element s drops below 40 25 C the visual alarm will automatically ca...

Page 8: ...ets disconnected during normal use the pad fault indicator and the 32oC LED will be lit The beeper will also sound Switch the unit off and reconnect the Heating Pad Restart the unit and observe a correct self test 4 7 CONTROL UNIT FAULT Any fault in the Control Unit will lit the Control Unit fault symbol and activate the acoustic alarm The heating will be switched off Switch the unit off and try a...

Page 9: ...y a technician 5 3 SELECTING TEMPERATURE In case you wish to select another temperature than 37 C then do as described in chapter 4 4 1 It will take about 8 15 minutes for the elements in the pad to reach the set temperature depending on surrounding conditions NOTE Please refer to section 2 regarding choice of temperature 5 4 GEL PADS Always use Gel Pads to get an even heat distribution a greater ...

Page 10: ... 1 2 MAINS POWER FAILURE ALARM AND RESTARTING Connect the Heating Pad to the Control Unit and the Control Unit to the mains Switch on the unit and wait for the self test to be completed Disconnect the power cord The alarm shall sound for at least 10 minutes Reconnect the power cord The unit shall perform the self test and then warm at the temperature that was selected before power off 6 1 3 FAULTS...

Page 11: ...ntact your local dealer 8 EMC COMPATIBILITY STATEMENT Please see www kanmed se section downloads Operatherm 9 WARRANTY STATEMENT KanMed warrants the purchaser that the Operatherm Control Unit and Heating Pads shall be free from defects in material and workmanship for a period of 12 month from the date of purchase The sole obligation of KanMed with respect to any such defect is limited to the repai...

Page 12: ...ried out by the user operator Self test at Power ON Every time the equipment is switched on Visual inspection of pad cable and connector Before every use Checks carried out by technical personnel Alarm functions Once a year Mains failure Pad failure Over temperature Temperature control Once a year Electrical safety Once a year Exchange of connector saver Once a year 10 3 STORING The Heating Pad sh...

Reviews: