User Manual
BW-50-070/10
Page 14 of 16
12 Technical data
Heating Pad
Voltage
24 Volt AC from Control Unit BW-50 when powering with mains power or 24 V DC
when powering from 24V battery source
Power consumption
50 Watt ,12 Watt when powering the Control Unit from a 12 Volt battery.
Dimensions and Weight
Length 580 mm, With 250 mm, Thickness 2 mm, Weight 0.5 Kg
Length of connection cable
1,2 m
Material cover
PVC
Watertight
Yes, except connector
Conductivity - surface
Not electrically conductive
Radiation
No measurable magnetic and electrical fields
Control Unit
Mains power
220 -240 V AC or 110 - 120 V AC
Battery power
12 - 24 V DC ( Limited warming capacity at 12 V DC)
Frequency
50 - 60 Hz
Power consumption
Max 65VA
Temperatures available
34
O
C - 38
O
C in steps of 0.5
O
C
Accuracy ±0.5
O
C
Resolution 0.5
O
C
Auditory Alarm
55 dBA Sound level
Indication of temperatures
and alarms
LED 35 -37
O
C are green, 34, 34.5, 37.5 and 38
O
C are yellow
< 34, >38
O
C are red alarms
Over temperature alarm
At a water temperature of 39.5
±
1
O
C
Dimensions and Weight
Length 210 mm, With 165 mm, Height 120 mm, Weight about 3 kg
Fuses F1, F2
220-240 V AC: T 1A L250V, 110-120 V AC: T 2A L250V
Mode of use
Designed for continuous use
Water protection
Drip proof when installed according to section 2
Internal Battery
Alkaline, 1.5V size LR6(AA) (Only for power failure alarm)
Water Mattress
Dimensions and Weight
Length 600 mm, With 270 mm, Height 30 mm,
Other sizes might be available. Please contact your local supplier.
Weight about 0.5 kg
Material PVC
Anti Algae recommended
KANMED Bottle 119 ml fluid
Poly [Oxyethylene (dimethylimino) ethylene (dimethyliminio)
ethelyenedichloride] 10% Inert ingredients 90%
KANMED Baby Nest
Dimensions
Length 650 mm, Width 450 mm Height 50 mm flattened
Material
Cover: High quality cotton and polyester
Filling of collar: Polyester fibre. Always tumble dry
Washing
Recommended temperature 60
0
C, Can tolerate 90
0
C but may shorten lifetime
Other sizes and materials are available. Please contact your local supplier.
Warming Capacity
(Fill with warm water to reduce the warming up time)
Normal conditions
About 4-5
0
C per hour. (Room temperature about 22
0
C and filled with 4,5 litres of
water and placed on the mattress of a baby bed and covered with the nest and a
sheet. Connection to mains or a 24V battery)
12 V Battery
About 1.5
0
C per hour up to set temperature minus 6
0
C
About 0.5
0
C per hour up to set temperature minus 2
0
C
About 0.3
0
C per hour up to set temperature
Twin Mattress
If not filled with warm water the
∆
T alarm will be activated after 4 hours. Restart and
the temperature will be reached. A “naked” twin mattress will not reach set
temperature due to excessive heat loss