SPECIFICATIONS
PARAMETER
RESOLUTION
ACCURACY
RANGE
Temperature & Pressure Measurement
Flue Temperature
Inlet Temperature
0.1°C
0.1°C
±0.1°C
±0.3% reading
±0.1°C
±0.3% reading
-
50
-
1200°C
With suitable probe
0
-
50°C
Pressure (Differential)
0.1mbar
±0.5% FSD full scale
±150mbar
Flue Gas Measurement *1
Oxygen
0.1%
±0.3% Volume
0
-
25%
Carbon Monoxide
High Range NDIR
0.1%
±5% or reading from
0.1% to 10%
0
-
10%
Hydrogen Sulphide
(optional)
1ppm
±5ppm<100ppm
±5%>100ppm
0
-
200ppm
Nitric Oxide
(optional)
1ppm
±5ppm<100ppm
±5%>100ppm
0
-
5000ppm
Nitrogen Dioxide
(optional)
1ppm
±5ppm<100ppm
±10ppm<500ppm
0
-
1000ppm
Sulphur Dioxide
(optional)
1ppm
±5ppm<100ppm
±5%>100ppm
0
-
5000ppm
Carbon Dioxide
NDIR
0.1%
±0.3% reading
0
-
20%
Hydrocarbon
NDIR
1ppm
+/
-
5% of reading or
+/
-
12ppm volume
0
-
5000ppm
Over
-
range:
10,000ppm
KANE988 MANUAL 39
Carbon Monoxide
(H2 Compensated)
1ppm
0
-
10000ppm
10000
-
20000ppm
±5ppm<100ppm
±5%>400ppm
-
2000ppm
±10%>2000ppm