USING YOUR ANALYSER
CHECK BEFORE SWITCH
-
ON:
Particle and water stop filter are dry and clean
Water trap and probe line are empty of water
Water trap is correctly fitted and instrument upright
All hoses connections, etc, are properly made
Flue temperature is connected
Analyser & probe will sample fresh outdoor air
Power ON instrument by pressing
to start automatic calibration count
down.
AUTOMATIC CALIBRATION
During automatic calibration analyser samples fresh air to zero toxic sensors
and set oxygen sensor to 20.95%.
After power on your analyser displays identity, software version and
serial number.
“
ANALYSER PURGING 90 secs
”
countdown appears on display.
Calibration time counts down in seconds to zero and can be changed to
90, 120, 180 or 300 seconds.
NOTE:
-
180 seconds is recommended to allow sensors to fully stabilise
-
anything less may result in toxic and oxygen sensor drift.
KANE988 MANUAL 21