13. COLD WEATHER PRECAUTIONS
It is important you keep your flue gas analyser in a warm place overnight
Electronic devices that become really cold, by being left in a vehicle
overnight, suffer when taken into a warm room the next morning.
Condensation may form which can affect the analyser’s performance &
cause permanent damage
Electrochemical sensors used in flue gas anlysers can be affected by
condensation or water being sucked into the analyser, as the small
apertures on top of sensors can become blocked with water, stopping
sensors seeing flue gas. When this happens, oxygen or carbon dioxide
reading will display as "—" & sensors may be permanently damaged
If you think that your analyser is affected by condensation or water
ingress, it may be possible to rectify the problem yourself. Simple leave
the analyser running in a warm place, with the pump 'ON' sampling fresh
air for a few hours (use mains adapter/battery charger if needed). If, after
doing this, you still experience problems please contact our Service
Centres.