2. BEFORE USING THE ANALYSER EVERY TIME:
Check the water trap is empty and the particle filter is not dirty:
To empty water trap, unscrew the red screw plug and re-tighten once it is
empty.
To change the filter, remove protective rubber cover, pull out the water
trap unit from the analyser, remove the water trap’s particle filter from its’
spigot and replace. Reconnect the water trap and protective rubber
cover.
Connect the flue probe hose to the analyser’s flue gas inlet and connect
the flue probe’s temperature connector to the T1 socket – check the
plug’s orientation is correct otherwise incorrect temperature
measurements will occur.
After switch on, check fuel source, date and time are correct and battery
power is sufficient.
3. FRESH AIR PURGE
Position the flue probe in fresh air, then press ON/OFF. The analyser’s
pump starts and the analyser auto calibrates for approximately 75
seconds.
Using the rotary dial select CO
2
& CO. In fresh air both COppm and
CO
2
% reading should be zero.
4. STATUS DISPLAY
Select the “Status” position on the dial to view Battery Status and
Calibration Due Date.