Kamoer fluid tech(Shanghai)Co., Ltd. UIP WIFI User Manual
13
Figure 2-4-1 System Settings
a. Tube type:
the current calibrated tube model;
b. Flow rate:
the flow rate before calibration;
c. Speed:
the speed corresponding to the current flow rate;
d. Calibration progress:
starting from 0, when the calibration is completed, it is 100;
e. Calibration time:
the length of time the calibration needs to be run, and the longer the
calibration time setting will make the calibration more accurate if the calibration container allows
it;
f. Calibration time:
click to set the calibration time;
g. Flow rate setting:
click to set the calibration flow rate, change the flow rate will change the
corresponding speed;
h. Prime:
Click to start priming, click again to stop emptying, when calibrating, you need to
empty the air inside the pump tube;
i. Start/Stop:
click start, the pump starts running, the calibration progress will show the progress.
When the calibration time countdown is completed, the pump stops running, the interface will pop
up the input box, and the volume obtained by the pump operation needs to be input to complete
the calibration;
Calibration process and areas to be aware of:
1. Preparation before calibration: Connect the pump tube outlet into the measuring cylinder, put
the water inlet into the water container, and fix the pump tube outlet and water inlet to prevent
the pump tube from sprinkling water when the pump works. Operation console
2. Enter the calibration function: in the system settings -> tube type settings -> select the tube
type that needs to be calibrated to enter the tube type calibration interface;
3. Set the calibration time: set the time as long as the cylinder allows, so that the accuracy of the
pump will be higher after calibration;
4. Flow rate setting: set a flow rate that needs to work;
5. Drain the pump tube: Click the empty button to let the water fill the pump tube and empty the