7
2.2 Ration mode
After the power is turned on, clicking the Ration mode in the title bar will enter the ration
mode work interface; in the ration mode, there are two working modes, namely the fully automatic
working mode and the semi-automatic working mode;
1. Semi-automatic working mode:
Click h working volume display area to set the volume.
After determining the quantity, click the start button to start running, the flow rate starts to count
in the middle of the screen. When the counting value and the setting amount are equal, the pump
stops working, click again to start. , the pump is again operated in the above manner;
2, fully automatic working mode:
set the number of cycles, set the working volume, pause
time, click to start running, the middle of the screen shows the flow starts counting, when the
count value and the set quantitative are equal, the pump stops running, and the middle of the
screen begins to The pause time countdown, when the countdown is 0, the pump automatically
restarts the work, the pump runs once and records it as a cycle. When the cycle is set a number of
times, the pump stops running, waiting for the click to start, and starts the next cycle;
Figure 2-2-1 Assignment mode
a, Ration mode:
allocation mode title;
b. Tube type:
the ration mode tube type setting, the tube type record has the corresponding flow
rate information, and the same speed will use different tube type flow rates;
c, Flow rate:
the current flow rate setting of the ration mode;
d, Speed:
the current set speed of the ration mode, the speed can not be directly set, the speed is
converted according to the current flow rate setting;
e, Temperature:
shows the temperature of the external temperature sensor, when the external
sensor is not connected, it is displayed here;
f, Working volume display:
display the currently set working volume;
g, Copy function entry:
When you do not know how much volume you need to add, you can use