Place a funnel in stopper hole at the top of the Cathode Solution Cell
(G)
and pour entire
contents of KAM
®
Generator Solution C (5 ml) ampoule into the Cathode Solution Cell
(G)
.
Remove funnel and place the glass Stopper
(H)
in place.
BASE ASSEMBLY
NOTE: Before assembling the base unit and glassware, ensure that the unit has been turned off.
9.
1.
2.
3.
4.
4. OPERATION continued
NOTE: Generator Solution C ampule is glass. To open, hold the ampule in a paper or cloth towel,
and apply pressure to the top of the ampule just above the small dot. Top should cleanly snap off
the base of the ampule.
Titration Cell Sleeve
Swivel Clamp
Cathode Solution
Cell Jack (black)
Detector Electrode
Jack (red)
Turn the swivel clamp toward back of KF unit and place the entire Generator Solution Cell
assembly into the Titration Cell Sleeve. Turn Swivel Clamp over the top of the cell and secure by
hand tightening the screw on top of the clamp. Do not over tighten.
Plug the free end of the Cathode Solution Cell Wire into the black colored jack on the side
panel of the base unit, and the free end of the Detector Electrode Wire into the red colored jack.
If using a portable unit, remove the power cord from packing or storage compartment and plug
into the power jack on the back of the base. Plug unit into power source and press power but-
ton.
The power button should glow blue. If you are using a portable unit, the light will blink when the
battery is charging. The LCD display should show the Main Menu.
Titration Cell Assembly