5. MAINTENANCE
REAGENT REPLACEMENT PROCEDURE
Unplug Detector Electrode and Cathode Cell wires from their respective jacks.
Loosen screw on the top of the swivel clamp and turn clamp toward back of unit.
Carefully remove entire glassware unit and place on a clean surface.
Remove Glass Stopper from the Cathode Solution Cell and wipe free of grease. Also
remove grease from opening at the top of the cell.
Remove the Cathode Solution Cell from the Generator Solution Cell by gently twisting and
lifting. DO NOT pull using the wire. Remove grease from the exterior of the Cathode Solu-
tion Cell and opening in the Generator Solution Cell opening.
Remove Detector Electrode (do not pull by wire), Sample Injection Port, and Drying Tube
from the Generator Solution Cell.
Remove the Glass Stopper from the Cathode Solution Cell.
Empty contents of Cathode Solution Cell by pouring into a suitable container or aspirating
from the cell. Dispose of fluids according to SDS sheet(s).
Empty contents of Generator Solution Cell by pouring into a suitable container or aspiration.
Dispose of fluids according to the SDS sheet.
Thoroughly wash the Generator Solution Cell with water and dish soap, then rinse with
water and dry by inserting a paper towel in large hole in lid and wiping down inside walls of
the cell.
Follow steps for Glassware Assembly and Base Assembly on pages 7-8 of this manual.
Follow steps to Reset Reagent Life on page 9 of this manual.
Follow steps for Initial Titration on page 10 of this manual.
Both time of use and total volume of samples affect re-
agent life. The KAM
®
KF unit automatically notifies you
when reagents need to be replaced. Current
reagent life is indicated on the Main Menu, and when
the reagents have expired a warning will appear on
the display for a few seconds between each sample
analysis until reagents have been replaced. To change
reagents, follow the steps below.
Refer to pages 7-8 for part images.
2.
1.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
!
R E A G E N T L I F E E XC E E D E D
P L E A S E R E P L A C E R E A G E N T
K a r l F i s c h e r M o i s t u r e A n a l y z e r