![KAM Simple Precision PKF User Manual Download Page 4](http://html1.mh-extra.com/html/kam/simple-precision-pkf/simple-precision-pkf_user-manual_3324268004.webp)
4
KFMANUAL 0317
KAM CONTROLS, INC.
Method:
Coulometric Karl Fischer titration
Detection:
Polarization detection
Control:
Automatic electrolysis current control
Display:
320 x 240 pixel LCD
Sample size:
0.1, 0.25, 0.5, 1.0 ml
or less than 2 grams (or ml)
Range:
10 μg - 100,000 μg H
2
O
Sensitivity:
1 μg H
2
O
Accuracy:
±5 μg for 10 μg - 1000 μg,
0.5% (C.V.) for over 1000 μg
(meets or exceeds API MPMS 10.9, ASTM D4928, ISO 12937,
and EI 386)
Generator electrode config.:
With diaphragm
Titration speed:
1000 μg H
2
O/min. (max. at high H
2
O concentrations)
Power requirements:
Operates on either AC or DC.
AC - 110/120, 220/240 V, 50/60 Hz
12 V rechargeable lithium battery
Ambient Temperature
5ºC – 40º C
Communication:
USB port, BlueTooth
®
, Windows
®
based software
Printer:
Optional
Dimensions:
Portable 15.5” x 8”x 10” (394 mm x 203 mm x 254 mm)
Lab 10.5” x 8” x 9”(267 mm x 203 mm x 229 mm)
Weight:
Portable
approx. 20 lbs. (9 kg)
Lab approx. 10 lbs. (4.5 kg)
2. SPECIFICATIONS