background image

2                                                                    Kalypto Medical  NPD 1000 140cc Dressing Application Guide 

 Kalypto Medical  NPD 1000 140cc Dressing Application Guide                    

                 3

Wound Site Preparation of the Foot for Dressing Application

Carefully inspect the wound and make sure no contraindications 

for NPWT are present 

(see the NPD 1000 System IFU P/N #50000 for 

a complete list of contraindications. If you do not have a copy, 

please see page 2 of this document for instructions to get one)

In preparation for applying the dressing, make sure proper hand 

washing and gloving practices are followed to prevent infection 

and cross-contamination. Clean the wound bed according to the 

institution’s protocol and standards of practice prior to application 

of the negative pressure dressing. To protect the peri-wound area 

(all skin that will be covered by the negative pressure dressing), 

care needs to be taken to adequately clean and dry this area. It is 

recommended that an alcohol-based cleaning product be used 

to prepare the peri-wound area, as this will effectively remove any 

remaining moisture. Make sure that any hair is clipped prior to the 

application of the dressing. A protective barrier wipe, such as Skin 

Prep

®*

, can be used to preserve the wound margins and prevent 

epithelial stripping. This will also insure that the proper adherence 

of the dressing to the site for the duration of wear.

REMOVE THE DRESSING FROM THE PACKAGE

 

Package can be opened by grasping the “nick” at the top corner 

of the package and pulling

.  

Remove dressing slowly and carefully 

from the sterile packaging.  (Fig. 2)

 

 

 

 

Open the package so the package 

sides are separated-keeping the bottom 

connected. This will expose the sterile inner 

package. Place the package silver side 

up and use this to set the dressing down to 

connect the tubing prior to application. To 

connect the tubing simply connect the >>> 

end to the plastic connection point on the 

dressing. (Fig. 3)  

 

Fig. 3:

  Use the inside of the dressing 

packaging as a sterile placemat  

to connect dressing tube prior to  

application

Fig. 2:

  Open dressing 

sterile packaging

Summary of Contents for 140cc Dressing

Page 1: ...K a l y p t o M e d i c a l NPD 1000TM Negative Pressure Wound Therapy DRESSING APPLICATION GUIDE 140cc 12x16cm Dressing...

Page 2: ...on on how to apply the Kalypto Medical 140cc Dressing for use with the NPD 1000 NPWT System The NPD 1000 Dressing Technology is a state of the art product that delivers the benefits of negative pressu...

Page 3: ...be used to prepare the peri wound area as this will effectively remove any remaining moisture Make sure that any hair is clipped prior to the application of the dressing A protective barrier wipe suc...

Page 4: ...ation process but DO NOT stretch the dressing cover at the edges as you apply it NOTE As the dressing cover is a semi occlusive polyurethane stretching may induce unintended skin trauma in the peri wo...

Page 5: ...the skin taking care to minimize folds and wrinkles while sealing the gasket for a proper air tight seal Fig 7 4 Run your finger along the perimeter of the gasket to secure the gasket material to the...

Page 6: ...y negative pressure dressing care should be taken to secure the pressure tubing between the dressing and the pump such that it does not become a tripping hazard Running the tubing inside the patient s...

Page 7: ...t two minutes If the leak icon does not disappear the problem is in the pigtail connector and the clinician should be notified of the system situation issue 3 If the pump and the pigtail are fine the...

Page 8: ...ghts MN 55120 Website www kalyptomedical com Email info kalyptomedical com Phone 651 289 0284 TOLL FREE 877 286 3740 Fax 651 289 0285 P N 50037 Rev B Kalypto Medical 10 09 KALYPTO MEDICAL CONTACT INFO...