Easidew Transmitter User’s Manual
Kahn Instruments
7
INSTALLATION
2.5.2 Transmitter Mounting - Direct Pipeline Connection
The transmitter may be directly mounted into a pipe or duct, as shown in
Figure 8
CAUTION: Do not mount the transmitter too close to the
bottom of a bend where any condensate in the pipeline
might collect and saturate the probe.
The pipe or duct will require a thread to match the transmitter body thread. Installation
dimensions are shown in
Figure 8. For circular pipework, to insure the integrity of a
gas tight seal, a mounting flange will be required on the pipe in order to provide a flat
surface to seal against.
The following procedure must be carried out by
competent personnel.
1.
Insure that the blue protective cover (and its desiccant capsule) has been
removed from the tip of the transmitter.
WARNING: Under no circumstances should the
sensor guard be handled with the fingers.
2.
Install a bonded seal (2) over the threaded part of the transmitter body.
3.
Screw the transmitter (3) into the pipe (1). Tighten enough to obtain a
gas tight seal. (Torque will depend upon the pipeline material.)
NOTE:
Do not overtighten or the thread on the pipe may be stripped.
Figure 8
Transmitter Mounting - Pipe or Duct
Summary of Contents for Easidew