Easidew Transmitter User’s Manual
6
Issue 9.1 December 2012
INSTALLATION
2.5.1 Transmitter Mounting - Sample Block (Optional)
The following procedure must be carried out by a
qualified installation technician.
To mount the transmitter into the sensor block (preferred method), proceed as follows,
refer to
Figure 7.
1.
Insure that the protective cover (2), and its desiccant capsule (2a), have
been removed from the tip of the transmitter.
2.
Install the bonded seal (4) over the threaded part of the transmitter body.
WARNING: Under no circumstances should the sensor
guard be handled with the fingers.
3.
Screw the transmitter (1) into the sample block (3) and tighten to a
minimum torque setting of 22.5 ft-lbs (30.5 Nm).
NOTE: Use the flats
of the hexagonal nut and not the sensor body.
4.
Attach the transmitter cable/connector assembly (6) to the plug located
on the base of the transmitter and tighten the retaining screw (7) (see
Section 2.3).
Figure 7
Transmitter Mounting - Sensor Block
Summary of Contents for Easidew
Page 20: ...http www kahn com...