![K-Patents PR-14038 Instructions Manual Download Page 8](http://html1.mh-extra.com/html/k-patents/pr-14038/pr-14038_instructions-manual_648507008.webp)
6
Cooling cover instruction manual
Document/Revision No.
IM-EN-CCPR43
/1.04
Effective: April 23, 2019
2.2 Connecting cooling water
Cooling water recommendations
Temperature
5-25°C (41-77°F)
Pressure
max. 6 bar (87 psi)
Flow
max. 0.25 l/min (0.7 GPM)
The flow can be quite a bit below maximum as long as the sensor internal temperature
stays below 60°C (140°F). K-Patents recommends that used water is directed to a drain.
The coolant connection is a 1/8" NPT female thread. Connect inlet to one and outlet to
the other connection, it doesn’t matter which way they are connected, because cooling
cover is symmetrical.
P4-033
1/8” NPT (F) Thread
Figure 2.3
Connecting the coolant hose