2 Installation
5
© Copyright K-Patents 2019. All rights reserved.
1.
Remove cover plate and detach cable connectors
from the circuit board
2.
Then remove circuit board, desiccant dryer
and O-ring from the cover plate
3.
Install new O-ring and desiccant dryer
supplied with the cooling cover to cooling cover and then attach circuit board
4.
Reattach cable connectors to circuit board
and insert cooling cover into Refractometer and attach with cover plate
Cover plate
T20
T10
Cover plate
(do not reinstall O-ring)
M4x25 Torx A2 DIN7895 screws (supplied with cooling cover) T20
Cooling cover
Cooling cover
O-ring
O-ring
Desiccant dryer
Desiccant dryer
Circuit board
Cable connectors
Circuit board
Cable connectors
P4-032