![K-Patents PR-14038 Instructions Manual Download Page 6](http://html1.mh-extra.com/html/k-patents/pr-14038/pr-14038_instructions-manual_648507006.webp)
4
Cooling cover instruction manual
Document/Revision No.
IM-EN-CCPR43
/1.04
Effective: April 23, 2019
3. Detach the cable connectors from the circuit board. Then unscrew the circuit board.
Remove the desiccant and O-ring, they are not needed any more.
4. Add the new O-ring and the new desiccant to the cooling cover (fold the edges of
the desiccant under the package to make it fit the slot) and screw on the circuit
board. Reattach the cable connectors to the circuit board.
5. Insert cooling cover into the refractometer, add cover plate and fasten with the new
longer screws. Note that there is no O-ring between the cover plate and the cooling
cover.
Summary of Contents for PR-14038
Page 11: ......