WARNING:

 Adjacent and/or Stacked Equipment 

The Product should not be used adjacent to or stacked with 
other  equipment.    If  adjacent  or  stacked  use  is  necessary, 
the Product should be observed to verify normal operation in 
the configuration in which it will be used.

 

 

CAUTION:

 Systems Statement

 

Equipment connected to the product must be certified to the 
respective  IEC  Standards  (i.e.  IEC  950  for  data  processing 
equipment  and  IEC  601-1  for  medical  equipment). 
Furthermore,  all  configurations  shall  comply  with  the  system 
standard  IEC  601-1-1.  Anybody  who  connects  additional 
equipment  to  the  signal  input  part  or  signal  output  part 
configures  a  medical  system,  and  is  therefore,  responsible 
that the system complies with the requirements of the system 
standard  IEC  601-1-1.  The  product  service  port  is  only 
intended  for  use  during  maintenance  by  authorized  service 
personnel. 
 

CAUTION:

 Case Cleaning Solutions

 

When disinfecting the case, use a non-oxidizing disinfectant, 
such  as  ammonium  salts  or  glutaraldehyde  based  cleaning 
solution, to avoid damage to the metal connectors. 
 
 

CAUTION:

 

Environment of use 

The  product  is  designed  for  indoor  use.  Operator  must 
confirm  that  the  environment  of  use  meets  the  required 
operating environmental specifications before using. 

 

CAUTION:

 Cold Environments

 

If the product is stored in an environment with a temperature 
below the operating temperature, the unit should be allowed 
to  warm  up  to  the  needed  operating  temperature  before 
using.

 

 

                                                                                               

 

17

 

Overall sensitivity at the distances 
200mm from the face of the transducer 

Doppler frequency:300

±

50Hz,Targe 

velocity: 10cm/s~40cm/s

 

90dB 

Spatial-peak temporal-peak acoustic 
pressure: 

<8.6kPa 

Output power: 

<26mW 

Effective area of the ultrasonic 
transducer active element: 

5.3cm

2

±

20% 

The acoustic coupling medium for 
normal use: 

ph :5.5~8, Acoustic 
impedance: 
1.5*10

6

~1.7*10

6

Pa

·

s/m 

AUDIO OUTPUT 

 

Audio Output Power: 

<0.2 W 

Audio out socket: 

Φ

3.5mm 

BATTERY   

 

Battery Voltage: 

9V 

Type: 

IEC6F22 9V alkaline 

5.2 MODE OF OPERATION 

Continuous operating 

5.3 PHYSICAL DIMENSIONS 

104mm (W) x 120mm (D) x 61mm (H),      Wt: 0.14kg (including battery) 

 

4.1 in, D 

4.7in, H 

2.4 in,                Wt: 31lbs (including battery)

 

 

5.4 ENVIRONMENTAL REQUIREMENTS 

OPERATING CONDITIONS 

Temperature:    5

°

C to 40

°

Humidity:    <80% RH, non-condensing 
Atmospheric pressure: 86kPa to 106kPa 

Summary of Contents for AngelSounds JPD-100S

Page 1: ...PER MEDICAL EQUIPMENT CO LTD 6th Floor No 6 Bldg Nanshui Industry Village 5th Industry Rd Shenzhen 518067 P R China Fetal Heart Detector JPD 100S INSTRUCTION MANUAL Manual Ver 1 0 Issuing Date 2009 11...

Page 2: ...tion book Without written authorization from Jumper Medical Equipment Co Ltd any individual or organization shall not copy modify or translate this book All contents described in this book are consist...

Page 3: ...bsorption and reflection from structures objects and people TABLE OF CONTENTS SECTION 1 INTRODUCTION 1 1 OVERVIEW 1 2 PRODUCT DESCRIPTION 1 1 3 OPERATING PRINCIPLE 1 SECTION 2 SAFETY GUIDANCE 2 2 1 IN...

Page 4: ...retically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in whi...

Page 5: ...ith the controls and how to use the product properly before operating the product CAUTION It should not be used in life supporting or life sustaining applications 1 2 PRODUCT DESCRIPTION The product i...

Page 6: ...l Heart Detector needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided for in the ACCOMPANYING DOCUMENTS Portable and mobile...

Page 7: ...eath CAUTION This alert identifies hazards that may cause minor personal injury product damage or property damage 2 5 SAFETY ALERT DESCRIPTIONS The following is a list of product safety alerts that ap...

Page 8: ...damage to the metal connectors CAUTION Environment of use The product is designed for indoor use Operator must confirm that the environment of use meets the required operating environmental specificat...

Page 9: ...P R China Tel 86 755 26696279 Fax 86 755 26852025 Web site http www jumper medical com Wellkang Ltd Suite B 29Harley Street LONDON W1G9QR U K 5 2 6 SYMBOL DESCRIPTIONS The following symbols may appear...

Page 10: ...leaning solution on the surface of the device Wipe the surface of sensor of transducer with 70 ethanol or alcohol self air dry or clean with a clean dry cloth 4 6 DISINFECTIONS Cleaning the unit surfa...

Page 11: ...ipe the transducer body with soft cloth to remove any remaining coupling gel Clean with soap only 4 5 4 Wring any excess moisture from the cloth before cleaning 4 5 5 Avoid pouring fluids on the devic...

Page 12: ...cables as described in the Section 5 Broken parts cracks damage or extreme wear broken or bent connectors and pins after bending and flexing the cable Replace cable if any abnormalities are found Exa...

Page 13: ...Section 5 for temperature specifications The product requires no calibration 4 3 VISUAL INSPECTION The product and its accessories should be carefully inspected prior to installation once every 12 mon...

Page 14: ...d Generally the site of heart of fetus is 1 3 below of navel line at its earlier stage it then moves upward with increasing of gestational period and the site of heart of fetus will be a little deviat...

Reviews: