wave  reaches  the  fatal  heart.  Then  it  is  converted  into  voltage.  This 
Doppler  signal  is  detected  and  demodulated  from  the  received  signal. 
And  the  Doppler  frequency  is  consistent  with  the  rhythm  of  the  fetal 
systole  and  diastole.  Once  cardiac  valves  vibrate  and  a  Doppler 
frequency  excursion  is  formed.  It  is  transmitted  an  output  signal  of 
cardiac  valves  vibrating,  and  it  is  sent  to  the  loudspeaker  for  getting  a 
rhythmical sound with the fetal heartbeat.

 

 

SECTION 2: SAFETY GUIDANCE 

 

2.1 INDICATIONS FOR USE 

The  product  is  normally  applied  to  fetus  above  9~12  weeks  growth, 
difference in pregnant mater. 
   

Listen to SFH: 

Operator  can  listen  to  the  sound  of  fetal  heartbeat  from  the 
headset.   

 

Audio record:

   

The  sound  of  fetal  heartbeat  can  be  recorded  by  a  recorder 
which is connected with the product. 
As  a  safety  advisement  that  can  only  be  connected  with  a 
recorder  complied  with  the  safety  requirements  of  IEC 
60601-1. 
 

2.2 CONTRAINDICATIONS FOR USE 

Normally none, as a particular case, please consult your doctor. 
 

2.3 NOTE FOR HOME USE 

  Please consult your doctor. 
 

2.4 SAFETY TERMS AND CONDITIONS 

The  signal  words  shown  below,  left,  identify  the  potential  hazard 
categories. The definition of each category is as follows: 

 

                                                                                               

 

19

 

Appendix A: EMC Information-Guidance and Manufacture

Declaration 

 

CAUTION:

 

Fetal Heart Detector

 needs special precautions regarding EMC 

and needs to be installed and put into service according to the 
EMC  information  provided  for  in  the  ACCOMPANYING 
DOCUMENTS. 
 
Portable  and  mobile  RF  communications  equipment  can  affect 

Fetal Heart Detector

 
The 

Fetal  Heart  Detector 

should  not  be  used  adjacent  to  or 

stacked with other equipment. 

 
The 

Fetal  Heart  Detector

  is  intended  for  use  in  the  electromagnetic 

environment specified below. The customer or the user of the Fetal heart 
monitor should assure that it is used in such an environment. 

Emissions 

 

test

 

Compliance

 

RF 
emissions 
CISPR 11 

Group 1 

The 

Fetal  Heart  Detector

 uses RF energy 

only  for  its  internal  function.  Therefore,  its 
RF  emissions  are  very  low,  and  are  not 
likely  to  cause  any  interference  in  nearby 
electronic equipment. 

RF 
emissions 
CISPR 11 

Class B 

The 

Fetal Heart Detector

 is suitable for use 

in  all  establishments,  including  domestic 
establishments 

and 

those 

directly 

connected  to  the  public  low-voltage  power 
supply network that supplies buildings used 
for domestic purposes. 

The 

Fetal  Heart  Detector

  is  intended  for  use  in  the  electromagnetic 

environment specified below. The customer or the user of the Fetal Heart 
Detector should assure that it is used in such an environment. 
Immunity test 

IEC 
60601 
test level 

Compliance 
level 

Electromagnetic 
environment guidance 

Electrostatic 
discharge 
(ESD) 

±

6 kV 

contact 

±

8 kV 

±

6 kV 

contact 

±

8 kV air 

Floors  should  be  wood, 
concrete or ceramic tile. If 
floors  are  covered  with 

Summary of Contents for AngelSounds JPD-100S

Page 1: ...PER MEDICAL EQUIPMENT CO LTD 6th Floor No 6 Bldg Nanshui Industry Village 5th Industry Rd Shenzhen 518067 P R China Fetal Heart Detector JPD 100S INSTRUCTION MANUAL Manual Ver 1 0 Issuing Date 2009 11...

Page 2: ...tion book Without written authorization from Jumper Medical Equipment Co Ltd any individual or organization shall not copy modify or translate this book All contents described in this book are consist...

Page 3: ...bsorption and reflection from structures objects and people TABLE OF CONTENTS SECTION 1 INTRODUCTION 1 1 OVERVIEW 1 2 PRODUCT DESCRIPTION 1 1 3 OPERATING PRINCIPLE 1 SECTION 2 SAFETY GUIDANCE 2 2 1 IN...

Page 4: ...retically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in whi...

Page 5: ...ith the controls and how to use the product properly before operating the product CAUTION It should not be used in life supporting or life sustaining applications 1 2 PRODUCT DESCRIPTION The product i...

Page 6: ...l Heart Detector needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided for in the ACCOMPANYING DOCUMENTS Portable and mobile...

Page 7: ...eath CAUTION This alert identifies hazards that may cause minor personal injury product damage or property damage 2 5 SAFETY ALERT DESCRIPTIONS The following is a list of product safety alerts that ap...

Page 8: ...damage to the metal connectors CAUTION Environment of use The product is designed for indoor use Operator must confirm that the environment of use meets the required operating environmental specificat...

Page 9: ...P R China Tel 86 755 26696279 Fax 86 755 26852025 Web site http www jumper medical com Wellkang Ltd Suite B 29Harley Street LONDON W1G9QR U K 5 2 6 SYMBOL DESCRIPTIONS The following symbols may appear...

Page 10: ...leaning solution on the surface of the device Wipe the surface of sensor of transducer with 70 ethanol or alcohol self air dry or clean with a clean dry cloth 4 6 DISINFECTIONS Cleaning the unit surfa...

Page 11: ...ipe the transducer body with soft cloth to remove any remaining coupling gel Clean with soap only 4 5 4 Wring any excess moisture from the cloth before cleaning 4 5 5 Avoid pouring fluids on the devic...

Page 12: ...cables as described in the Section 5 Broken parts cracks damage or extreme wear broken or bent connectors and pins after bending and flexing the cable Replace cable if any abnormalities are found Exa...

Page 13: ...Section 5 for temperature specifications The product requires no calibration 4 3 VISUAL INSPECTION The product and its accessories should be carefully inspected prior to installation once every 12 mon...

Page 14: ...d Generally the site of heart of fetus is 1 3 below of navel line at its earlier stage it then moves upward with increasing of gestational period and the site of heart of fetus will be a little deviat...