![JUMO Wtrans T03 Operating Manual Download Page 36](http://html1.mh-extra.com/html/jumo/wtrans-t03/wtrans-t03_operating-manual_2031193036.webp)
5 Annex
36
Environmental influences
Ambient temperature range of the
housing
-30 to +85 °C (housing incl. electronics)
If the operating temperature exceeds the max. ambient temperature, take into
account the heat input from the protection fitting into the housing, see
chapter 1.10 "Heat input through protection tube / protection tube fitting",
Page 12.
Storage temperature range
-40 to +85 °C (housing incl. electronics)
Temperature influence
≤ ±0.0025 %
a
/K;
per K deviation from the reference temperature 22 °C (±3 K) (housing)
Site altitude
Max. 2000 m above sea level
Resistance to climatic conditions
Rel. humidity ≤ 95 % without condensation according to IEC 68-2-30
Vibration resistance
Max. 2 g at 10 to 2000 Hz (relating to the housing including electronics)
according to DIN IEC 60068-2-6
Admissible mechanical
shock resistance
25 g for 6 ms (relating to the housing including electronics)
DIN IEC 68-2.29 per 1000 cycles
EMC
- Interference emission
- Interference immunity
- Radio frequency spectrum
DIN EN 61326-1
Class B - household and small businesses -
Industrial requirements
ETSI EN 300 220-1, ETSI EN 300 220-2
a
All accuracy specifications in % with regard to the measuring range of 290 K.
Housing
Material
Basic housing: PA 6, blue
Housingcover: PA 6.6 ESD, black
Flammability class
UL 94 HB
Dimensions
Diameter approx. 32 mm, length approx. 126 mm;
for insertion lengths refer to order details
Mounting
Can be used inside and outside of buildings
Protection type
IP67 according to DIN EN 60529;
for basic type 902930/15 and 902930/17;
for basic type 902930/55 only with machine connector M12 × 1 connected
Cleaning
The device can be cleaned with conventional washing, rinsing, and cleaning
agents
Installation position
Any
Weight
Approx. 120 g (without protection fitting)
Ex identification marking
Zone
Identification marking
Use in zone 0 (20)
TÜV 16 ATEX 177609 X
II 1 G Ex ia IIB T4 Ga
II 1 D Ex ia IIIB T130 °C Da
-30 °C ≤ Ta ≤ +85 °C
Use in zone 0 (20) / 1 (21)
with zone separation by protection
fitting
TÜV 16 ATEX 177609 X
II 1/2 G Ex ia IIB T4 Ga/Gb
II 1/2 D Ex ia IIIB T130 °C Da/Db
-30 °C ≤ Ta ≤ +85 °C
Summary of Contents for Wtrans T03
Page 2: ......
Page 4: ...Contents ...
Page 14: ...1 Introduction 14 ...
Page 24: ...2 Identifying the device version 24 ...
Page 30: ...3 Preparing the transmitter 30 ...
Page 34: ...4 Transmitter range 34 ...
Page 43: ...43 5 Annex 5 3 Examination certificate ...
Page 44: ...5 Annex 44 ...
Page 45: ...45 5 Annex ...
Page 46: ...5 Annex 46 ...
Page 47: ......