![JUMO Wtrans T03 Operating Manual Download Page 35](http://html1.mh-extra.com/html/jumo/wtrans-t03/wtrans-t03_operating-manual_2031193035.webp)
35
5 Annex
5
Annex
5.1
Technical data
Analog input
Measurement input
Pt1000 according to DIN EN 60751, in three-wire circuit
Operating temperature range
-30 to +260 °C
Accuracy
of the temperature sensor
Class A ±(0.15 K + 0.002 × ItI)
ItI = measured temperature in °C without prefix sign
Sensor line resistance
≤ 11 ohm per line for three-wire circuit
Sensor current
≤ 500 µA
Measuring circuit monitoring
Detection of probe break and probe short circuit
Output (wireless transmission)
Transmitter detection (transmitter
ID)
Max. 5-digit ID, factory-set, not configurable
Transmission interval
Adjustable via DIP switch: 5 s, 10 s, 20 s, or 45 s
(default setting = 20 s)
Radio frequency
868.4 MHz
Transmission power
< +10 dBm
Open air range
300 m and use of receiver
antenna holder for wall mounting and 3 m antenna cable.
When mounting the antenna directly to the receiver, a reduction in the range of
approx. 40 % must be taken into account.
Output signal
882.2 to 1977.1 ohm = -30 to +260 °C
Response time
of the complete probe
0.4 m/s in water; 3.0 m/s in air
Dia. 4.0 mm:
Water t
0.5
approx. 3 s, t
0.9
approx. 7 s; air t
0.5
approx. 25 s, t
0.9
approx. 80 s
Dia. 6.0 mm:
Water t
0.5
approx. 4 s, t
0.9
approx. 10 s; air t
0.5
approx. 32 s, t
0.9
approx. 98 s
Calibration accuracy
of electronics
≤ ±0.05 %
a
Voltage supply
Lithium battery
Voltage: 3.6 V, rated capacity: 2.2 Ah (
Symbol for direct voltage)
Operating life
Approx. 1 year with the factory-set values and at room temperature (a fast
transmission interval and a high or low ambient temperature reduce the battery
operating life)
Battery change
Only use the preconfigured lithium battery that is available as an accessory
(part no. 00525539)
a
All accuracy specifications in % with regard to the measuring range of 290 K.
Summary of Contents for Wtrans T03
Page 2: ......
Page 4: ...Contents ...
Page 14: ...1 Introduction 14 ...
Page 24: ...2 Identifying the device version 24 ...
Page 30: ...3 Preparing the transmitter 30 ...
Page 34: ...4 Transmitter range 34 ...
Page 43: ...43 5 Annex 5 3 Examination certificate ...
Page 44: ...5 Annex 44 ...
Page 45: ...45 5 Annex ...
Page 46: ...5 Annex 46 ...
Page 47: ......