6 Device settings
26
Parameter "t.rEF": setting the reference temperature
The parameter "t.rEF" can only be accessed if the setting option "Lin" or "nLF"
has been chosen when selecting the temperature compensation (t.Cor).
✱
Use the
and
keys to set the reference temperature:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Parameter "C.int": calibration timer
✱
Use the
and
keys to set the calibration timer:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Menu/parameter
Setting option
Meaning
25°C/77°F
Reference temperature 25°C/77°F
20°C/68°F
Reference temperature 20°C/68°F
Menu/parameter
Setting option
Meaning
1 to 730
Time interval for calibration timer in days
oFF
No calibration timer