6 Device settings
24
Parameter "rEF.S": selection of the standard reference solution for the calibration of the rel-
ative cell constant
If a selection has been made to calibrate the relative cell constant with a selec-
tion of standard reference solutions (see page 23), use this parameter to spec-
ify a standard reference solution for calibrating the relative cell constant.
✱
Use the
and
keys to select one of the standard reference solutions:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Parameter "Unit t": temperature unit °C/°F
✱
Use the
and
keys to select the temperature unit:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Menu/parameter
Setting option
Meaning
1413 µS
Reference solution 0.01 M KCl
2760 µS
Reference solution 0.02 M KCl
12.88 mS/cm
Reference solution 0.1 M KCl
50 mS/cm
Seawater comparative solution KCl
111.8 mS/cm
Reference solution 1 M KCl
Menu/parameter
Setting option
Meaning
°C
All temperatures stated in degrees
Celsius
°F
All temperatures stated in degrees
Fahrenheit