5.1
Display and control elements
Once the voltage supply has been applied, the LED initially illuminates orange for approx. 2 seconds, then green.
To check the correct function, it is best to connect the device to a PC using the setup program.
v
In the process data window, the current measured value, the output, device status, and the terminal temperature are displayed.
Caution
As the measurement input and the USB interface are not galvanically isolated, in the event of a grounded sensor, a
ground connection with a PC which is also grounded (e.g., with desktop computers) via USB must be avoided.
Please check the isolation on the sensor side, or use a laptop in battery mode for setup applications.
Transfer of the setup data is also possible even without a voltage supply to the device (power supply via USB).
In this case, only the processor and the analog input are supplied. No signal is issued at the output.
Summary of Contents for 00718766
Page 4: ...Inhaltsverzeichnis 4...
Page 10: ...3 Montage 10 3 Montage 3 1 Abmessungen...
Page 22: ...5 Ger t in Betrieb nehmen 22...
Page 35: ...9 China RoHS 35...
Page 40: ...table of contents 4...
Page 46: ...3 Mounting 10 3 Mounting 3 1 Dimensions...
Page 58: ...5 Startup of the device 22...
Page 71: ...9 China RoHS 35...
Page 76: ...Sommaire 4...
Page 82: ...3 Montage 10 3 Montage 3 1 Dimensions...
Page 94: ...5 Mise en service de l appareil 22...
Page 107: ...9 China RoHS 35...
Page 112: ...Tabla de contenido 4...
Page 118: ...3 Montaje 10 3 Montaje 3 1 Dimensiones...
Page 130: ...5 Poner en marcha el dispositivo 22...
Page 143: ...9 China RoHS 35...