7
、
OTHER ATTENTIVE ITEMS
35
Recommended separation distances between portable and mobile RF
communication equipment and the
JMC5A Ni Oxygen Concentrator
.
The
JMC5A Ni Oxygen Concentrato
r is intended for use in an electromagnetic
environment in which radiated RF disturbances are control . The customer or user
of the
JMC5A Ni Oxygen Concentrato
r can help to help prevent electromagnetic
interferences by maintaining minimum distances between the portable andmobile
RF communication equipment (transmitters) and the
JMC5A Ni Oxygen
Concentrato
r as recommended below, according to the maximum output power of
the communication equipment.
Rated maximum
output power of
transmitter
(W)
separation distance according to frequency of
transmitter(m)
150 kHz to 80 MHz
d=1.2√P
80 MHz to 800
MHz
d=1.2√P
800 MHz to
2.5 GHz
d=2.3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power is not listed above, the
recommended separation distance in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is themaximum
output power rating of the transmitter in watts (w) according to the transmitter
manufacturer. Specified by the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higherfrequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objectsand
people.