Julabo FD200 Operating Manual Download Page 3

Operator responsibility – Safety instructions 

14 

 

Environmental conditions according to IEC 61 010-1: 

 

Use indoors only. 

 

Altitude up to 2000 m - normal zero. 

 

Ambient temperature: see Technical specifications  

 

Humidity: 

 

Max. relative humidity 80% for temperatures up to +31 °C, 

 

linear decrease down to 50% relative humidity at a temperature of +40 °C 

 

Max. mains voltage fluctuations of ±10% are permissible. 

 

Protection class according to IEC 60 529 

IP21 

The unit corresponds to Class I 

Overvoltage category  

II 

Pollution degree  

 

 

Caution: 

The unit is not for use in explosive environment. 

 
EMC requirements according to EN 61326-1 
 

Information about the used refrigerants 

The

  Regulation (EU) No. 517/2014 on fluorinated greenhouse gases 

applies to all 

systems which contain fluorinated refrigerants and replaces (EC) 842/2006. 
The aim of the Regulation is to protect the environment by reducing emissions of fluorinated 
greenhouse gases. 
Among other things it regulates the emission limits, use and recovery of these substances. It 
also contains requirements for operators of systems which require / contain these 
substances to function. 
 
Under Regulation 517/2014, the operator of a system of this nature has the following duties: 
 

  The operator must ensure that the equipment is checked at regular intervals for leaks. 

 

  These intervals depend on the CO

2

  equivalent  of the system. This is calculated from 

the refrigerant fill volume and type of refrigerant. The CO

2

 equivalent of your system is 

shown on the model plate. 
 

  The operator undertakes to have maintenance, repair, service, recovery and recycling 

work carried out by certified personnel who have been authorized by JULABO. 

 

  All such work must be documented. The operator must keep records and archive them 

for at least five years. The records must be submitted to the relevant authority on 
request. 

 
Refer to the text of the Regulation for further information. 
 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for FD200

Page 1: ...02 05 2016 English Operating manual Immersion Coolers FT200 FT400 FT900 Flow Through Cooler FD200 FT900 FT 400 1 951 4616 V4 05 16 www GlobalTestSupply com Find Quality Products Online at sales Globa...

Page 2: ...5 V A 3 0 3 0 FT400 FT900 Temperature range C 40 30 90 30 Cooling capacity medium ethanol C kW 20 10 20 40 0 45 0 36 0 14 0 03 20 10 40 80 0 3 0 27 0 2 0 07 Cooling compressor 1 stage 2 stage Refriger...

Page 3: ...fluorinated greenhouse gases Among other things it regulates the emission limits use and recovery of these substances It also contains requirements for operators of systems which require contain thes...

Page 4: ...switch illuminated 2 Cooling control light 3 Clamp for immersion probe 7 Removable ventilation grid 8 Castor with locking lever 9 Tube connection discharge 10 Tube connection intake 11 Mains power cab...

Page 5: ...OPERATING CONTROLS AND FUNCTIONAL ELEMENTS 16 FT900 FT900 1 2 3 7 8 Rear view 11 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 6: ...ly dangerous situation If these instructions are not followed serious injury and danger to life could result Caution Describes a possibly dangerous situation If this is not avoided slight or minor inj...

Page 7: ...he air intake and exhaust grids free of obstructions Maintain a sufficient distance from all surrounding surfaces Do not move the unit from the position where it was set up during operation Always tur...

Page 8: ...R404A 1 m 3 of space is required For 0 68 kg of refrigerant R23 1 m3 of space is required The ambient temperature must not exceed 35 C Keep at least 20 cm of open space on the front and rear venting...

Page 9: ...tive earth The power supply plug serves as safe disconnecting device from the line and must be always easily accessible Never operate equipment with damaged mains power cables Regularly check the main...

Page 10: ...ate The immersion probe as part of the cooling circuit should not be exposed to bath temperatures above the working temperature of the immersion cooler This would cause damage to the compressor Do not...

Page 11: ...ays turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures Service and repair work may be performed only by authorized electrici...

Page 12: ...returning a JULABO instrument for repair please contact an authorized JULABO service station When returning the unit Clean the unit in order to avoid any harm to the service personnel Attach a short f...

Reviews: