19514616-V4.doc
02.05.2016
English
Operating manual
Immersion Coolers
FT200
FT400
FT900
Flow-Through Cooler
FD200
FT 400
1.951.4616-V4
05/16
www.
GlobalTestSupply
.com
Find Quality Products Online at:
[email protected]
Page 1: ...02 05 2016 English Operating manual Immersion Coolers FT200 FT400 FT900 Flow Through Cooler FD200 FT900 FT 400 1 951 4616 V4 05 16 www GlobalTestSupply com Find Quality Products Online at sales Globa...
Page 2: ...5 V A 3 0 3 0 FT400 FT900 Temperature range C 40 30 90 30 Cooling capacity medium ethanol C kW 20 10 20 40 0 45 0 36 0 14 0 03 20 10 40 80 0 3 0 27 0 2 0 07 Cooling compressor 1 stage 2 stage Refriger...
Page 3: ...fluorinated greenhouse gases Among other things it regulates the emission limits use and recovery of these substances It also contains requirements for operators of systems which require contain thes...
Page 4: ...switch illuminated 2 Cooling control light 3 Clamp for immersion probe 7 Removable ventilation grid 8 Castor with locking lever 9 Tube connection discharge 10 Tube connection intake 11 Mains power cab...
Page 5: ...OPERATING CONTROLS AND FUNCTIONAL ELEMENTS 16 FT900 FT900 1 2 3 7 8 Rear view 11 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 6: ...ly dangerous situation If these instructions are not followed serious injury and danger to life could result Caution Describes a possibly dangerous situation If this is not avoided slight or minor inj...
Page 7: ...he air intake and exhaust grids free of obstructions Maintain a sufficient distance from all surrounding surfaces Do not move the unit from the position where it was set up during operation Always tur...
Page 8: ...R404A 1 m 3 of space is required For 0 68 kg of refrigerant R23 1 m3 of space is required The ambient temperature must not exceed 35 C Keep at least 20 cm of open space on the front and rear venting...
Page 9: ...tive earth The power supply plug serves as safe disconnecting device from the line and must be always easily accessible Never operate equipment with damaged mains power cables Regularly check the main...
Page 10: ...ate The immersion probe as part of the cooling circuit should not be exposed to bath temperatures above the working temperature of the immersion cooler This would cause damage to the compressor Do not...
Page 11: ...ays turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures Service and repair work may be performed only by authorized electrici...
Page 12: ...returning a JULABO instrument for repair please contact an authorized JULABO service station When returning the unit Clean the unit in order to avoid any harm to the service personnel Attach a short f...