Menu functions
40
10.3. Adjustable
parameters
ACTUAL
SET
Set the parameters for the following menu items in the
configuration level:
OP
- Operating mode
0
=
Keypad
control
1 = Remote control via RS 232
ACTUAL
SET
HAnd
- Handshake of the serial interface
0 = XOn/XOff, software handshake
1 = RTS/CTS, hardware handshake *
ACTUAL
SET
PAr
- Parity bits of the serial interface
0
=
no
1
=
odd
2
=
even
*
ACTUAL
SET
bAud
- Baud rate of the serial interface
1200
Bauds
2400
Bauds
4800 Bauds*
9600
Bauds
ACTUAL
SET
E_Sb
- External stand-by for emergency switch-off
0 = stand-by input is ignored *
1 = stand-by input is active
(Stand-by input - see page 35)
(* factory setting)
ACTUAL
SET
Pr-t -
Programmer input type
0
=
Voltage
0 to 10 V
1
=
Current
0 to 24 mA