Manual operation
30
With the parameters HL (High Limit) and LL (Low
Limit) the temperature of the internal bath is limited
in case of external control. So, especially for big
consumers, a great overshoot resp. undershoot of
the internal temperature is avoided.
Effect of the limitation of the internal bath
temperature:
Without limitation
Limited internal temperature
Each indicated control parameter can be optimized
manually .
Setting
:
n
Operate the key
as often until the desired
control parameter is indicated. Example: CP3
o
Use the cursor keys
to move left or
right on the display until the numeral you wish
to change is blinking.
p
Use the increase/decrease arrows
to
change the selected numeral (-, 0, 1, 2, 3, ... 9).
q
Press enter
to store the selected value