Recirculating coolers
37
Command
Parameter Response of recirculating cooler
out_mode_05
out_mode_05
in_mode_05
0
1
no
Stop the recirculating cooler = rOFF
Start the recirculating cooler
Ask for actual condition (Start/Stop)
out_sp_00
in_sp_00
xx.x
no
Set working temperature value
Ask for working temperature value
in_sp_01
no
Ask for high temperature value
in_sp_02
no
Ask for low temperature value
out_sp_03
in_sp_03
xxx
no
Set control ratio for feed/return flow temperature
Ask for actual control ratio
out_sp_07
in_sp_07
xx.xx
xx.xx
Maximum value of the internal temperature in case of
external control (HL - High Limit)
out = set; in = ask
out_sp_08
in_sp_08
xx.xx
xx.xx
Minimum of the internal temperature in case of
external control (LL - Low Limit)
out = set; in = ask
in_pv_00
no
Ask for actual feed temperature
in_pv_01
no
Ask for actual temperature of external sensor
in_pv_02
no
Ask for actual heater capacity
in_pv_03
no
Ask for actual return temperature
in_pv_04
no
Ask for actual safety temperature
out_par_06
in_par_06
x:x
x:x
Control parameter Xp of the internal controller
out = set; in = ask
out_par_07
in_par_07
xx.xx
xx.xx
Control parameter Tn of the internal controller
out = set; in = ask
out_par_08
in_par_08
xx.xx
xx.xx
Control parameter Tv of the internal controller
out = set; in = ask
out_par_09
in_par_09
xx.xx
xx.xx
Control parameter Xp of the external controller
out = set; in = ask
out_par_10
in_par_10
xx.xx
xx.xx
Control parameter Xpu of the cascaded P-controller in
case of external control
out = set; in = ask
out_par_11
in_par_11
xx.xx
xx.xx
Control parameter Tn of the external controller
out = set; in = ask
out_par_12
in_par_12
xx.xx
xx.xx
Control parameter Tv of the external controller
out = set, in = ask