Operating procedures
6. Operating procedures
6.1. Power connection
Caution:
•
Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
•
The power supply plug serves as safe disconnecting device from the line
and must be always easily accessible.
•
Never operate equipment with damaged mains power cables.
•
Regularly check the mains power cables for material defects (e.g. for
cracks).
•
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply
voltage specified on the identification plate.
•
Connect the circulator with mains power cable (10a) to the
mains outlet (17).
•
Connect the refrigerated circulator with mains power cable
(10b) to the mains socket.
6.2. Switching on / Start – Stop
•
Switching on:
Circulator and cooling machine may be turned on and off with
separate mains switches. The integrated control light will
illuminate to indicate that power has been applied.
•
The unit performs a self-test. The segments of the 4-digit LED
temperature DISPLAY and all indicator lights will illuminate
(as
illustrated on the left).
Then the software version (example: 11.0) appears. The
display „
OFF
“ indicates the unit is ready to operate. (standby
mode).
Start:
•
Press the
key
for about 4 seconds.
The LED temperature DISPLAY indicates the actual bath
temperature.
Stop:
•
Press the
key for about 4 seconds.
•
Turn the unit off with the mains power switch.
26