Product description
14
5.4 Technical data
Performance specifications measured in accordance with DIN12876. Cooling
capacities up to 20°C measured with ethanol; over 20°C with thermal oil unless
specified otherwise. Performance specifications apply at an ambient
temperature of 20°C. Performance values may differ with other bath fluids.
Grouping of the device acc. to CISPR 11:
• The device is an ISM device of group 1, which uses high frequency for
internal purposes
• Class A: Use in an industrial electromagnetic environment
• The device is designed for connection to a separate power transformer or
generator. Connection to a low voltage line is not permitted
In accordance with IEC 61010-1, the device is designed for safe operation under
the following ambient conditions:
• Indoor use
• Altitude up to 2000 m above sea level
• Ambient tempe5 … +40°C
• Maximum relative humidity 80% for temperatures up to 31°C, decreasing
linearly down to 50% relative humidity at 40°C
• Mains voltage fluctuations up to ±10% of the nominal voltage permissible
if not otherwise specified
• Contamination level 2
Protection class according to EN 60 529:
• Protection class IP21
The technical data for the respective device combination can be found in the
operating manual of the associated circulator.