Troubleshooting

Troubleshooting

In Forehead mode:
Temperature taken is 
higher than 43.0 

(109.4

).

In Object mode:
Temperature taken is 
higher than 100 

(212

).

Problem

Problem

Solution

Solution

The thermometer could 
not

 

work due to low 

battery.

Replace two new 
alkaline batteries 
size AAA.
 

Error message              

Error message              

22

In Forehead mode:
Temperature taken is 
lower than 34.0 

(93.2

).

In Object mode:
Temperature taken is 
lower than 0

(32

).

21

The thermometer is not 
functioning properly.

Unload the battery, 
wait for 1 minute and 
repower it. If the 
message reappears, 
contact the retailer 
for service.
 

The ambient temperature
is not within the range 
between 10

 and 40

(50

~

104

).

Place the 
in a room for at least 30
minutes at room 
temperature between 
10

 and 40

(50

~

104

)

thermometer

Read Temperature Taking 
Hints Thoroughly, then 
take a new temperature
measurement. 

Read Temperature 
Taking Hints thoroughly ,
then make sure the lens 
filter are clean, then take 
a new temperature 
measurement. 

19

20

Battery Replacement

1.  Replace battery when "           appears in the upper right corner 

"

     of  LCD display. (See Figure 9)

2.  Slide battery cover down as shown in Figure 10.

3.  Remove battery and install 2 new AAA alkaline batteries 

     as shown 

 . 

in Figure 11

4.  Slide battery cover back on.

Figure 9

Figure 10

Figure 11

Specifications

Forehead mode: 34.0

~43.0

(93.2

~109.4

)

Object mode: 0

~100

(32

~212

)

Forehead mode: 

±

0.2

 (0.4

) during 35.5

~42.0

 (95.9

~107.6

at 15

~35

 (59.0

~95.0

) operating temperature range

±

0.3

 (0.5

) for other measuring and operating 

temperature range
Object mode: 

±

4% or 

±

2

(4

) whichever is greater

10

~40

(50

~104

), 15%~85%RH, non-condensing

700hPa ~ 1060hPa

Atmospheric Pressure : 

153*41*44mm

Approx.84grams(with batteries)

DC3V(2

×

AAA

 

battery)

Measuring range

Laboratory accuracy

Display resolution

0.1

 or 0.1

Measure time

Approximately 1 second

Operating temperature 
range:

Storage and transport 
temperature range

-25

~ 55

 (-13

~131

), 15%~95%RH, non-condensing

700hPa ~ 1060hPa

Atmospheric Pressure : 

Clinical accuracy

Dimension

Weight

Battery

Battery life

Approx. 1 year/6000 readings 

Measuring site    

        

Forehead(Forehead Mode)

 Oral (This thermometer converts the forehead temperature
 to display its “oral equivalent.”)

Reference body site

Operation mode                  Forehead mode(Adjust mode)

Ingress protecting rating

         
    

 IP22

Shock

withstands drop of 3 feet

Expected service life

Three years 

Clinical bias: -0.09

℃(-

0.16

℉)

Clinical repeatability: 0.13

℃(

0.23

℉)

Limits of agreement: 0.87

℃(

1.57

℉)

Summary of Contents for DET-306b

Page 1: ...itute for consultation with your physician 4 some parts are small enough to be swallowed 5 Never immerse this device in water or other liquids not waterproof 6 Do not modify this equipment without authorization of manufacturer 7 Do not expose the thermometer to temperature extremes below 25 13 or over 55 131 nor excessive humidity 95 RH 8 Keep the battery away from children 1 There is no gender or...

Page 2: ... For Use The real time clock will be recorded with the memory function and help you to recognize each measurement result Please see the Real time clock setting section to learn how to setup the time in the first use The thermometer has been designed for practical use It s not meant to replace a visit to the doctor Please also remember to compare the measurement result to your regular body temperat...

Page 3: ...l This is the only way to accurately diagnose a fever To determine normal temperature take multiple readings when healthy Re measure with a standard digital thermometer for confirmation 2 Users 3 Oils or cosmetics on the forehead may give a lower temperature reading than the actual one must be inside for 30 minutes before taking a measurement Note Users and the thermometer should be in the same am...

Page 4: ... a reading equal to or higher than 37 8 100 0 In Object mode The display will only be lighted GREEN for 3 seconds when the unit is ready for measurement and a measurement is completed 1 Press the BUTTON The display is activated to show all segments 3 Read the temperature on the display 4 Press the BUTTON to turn off After self checking Figure 2 appears on the display screen with beeps so you can s...

Page 5: ...to the EMC information Do not put the thermometer into water directly 7 Put the thermometer back to the original packaging after using 15 Match your with a Smart Device 1 If this is your first time using it bind first Open SETTING menu choose Bind and unbind device and select the appropriate model The date and time on your thermometer will automatically be updated when it s connected with your pho...

Page 6: ...ry cover down as shown in Figure 10 3 Remove battery and install 2 new AAA alkaline batteries as shown in Figure 11 4 Slide battery cover back on Figure 9 Figure 10 Figure 11 Specifications Forehead mode 34 0 43 0 93 2 109 4 Object mode 0 100 32 212 Forehead mode 0 2 0 4 during 35 5 42 0 95 9 107 6 at 15 35 59 0 95 0 operating temperature range 0 3 0 5 for other measuring and operating temperature...

Page 7: ...vice to the dealers or manufacturer The above recommendations do not supersede the legal requirements The user must always comply with legal requirements for the control of the measurement functionality and accuracy of the device which are required by the scope of relevant laws directives or ordinances where the device is used A clinical summary and procedures for checking calibration are availabl...

Page 8: ...e Use of the unit in conjunction with non approved accessories can affect the device negatively and alter the electromagnetic compatibility The device should not be used directly adjacent to or between other electrical equipment Electromagnetic Compatibility Information This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cau...

Page 9: ...d in the frequency range 80 MHz to 2 7 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause interference if it is inadvertently brought into patient areas For this reason an additional factor of 10 3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges c Field stre...

Page 10: ...or radio control products If you have to do so the device should be observed to verify normal operation The use of accessories and power cord other than those specified with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components may result in increased emissions or decreased immunity of the equipment or system ...

Reviews: