Joyfay Delicate BX80 Operating Instructions Manual Download Page 4

 

 

Detection mode:

 Gaseous diffusion   

Battery:

 Lithium ion rechargeable battery, 3.7V

 

Battery working period: 

About 200 hours 

Sensor work principle:

 Electrochemistry mode 

Sensor service life

: 2 years 

Display:

 Large screen LCD

 

Alarm:

 Audible and visible alarm

 

Direct readings:

 instantaneous value, peak value, battery voltage, TWA, STEL

 

Ex-mark:

 Ex ibdIICT4

 

Protection degree:

 IP45

 

Work temperature:

 -10~50

 

Work humidity:

 5~90% RH

 

Size:

 126mm (L) x 66mm (W) x 32mm (H)

 

Weight:

 220g (with charger)

 

3. Operating instruction 

 

 

Menu Key 

Power on/off Key 

Up Key 

Down Key 

Enter Menu 

Switch-on

 

Selecting menu

 

Selecting menu

 

Exit 

Switch-off

 

Value up

 

Value down

 

 

Selfcheck

 

Activate The Background 

light

 

Activate the background 

light

 

 

Zero Point Calibrating

 

 

 

 

Sensitivity Setting

 

 

 

 

Affirm Setting

 

 

 

Menu Key

 

Up Key

 

Down Key

 

Power on/off Key

 

Summary of Contents for BX80

Page 1: ...Portable O3 Ozone Tester Detector BX80 Operating instruction manual...

Page 2: ...ontent 1 General 3 2 Technical performance and parameters 3 3 Operating instruction 4 4 Adjusting and calibrating 11 5 Daily use and maintenance 14 6 Storage and quality assurance 14 7 Complete set 15...

Page 3: ...BX80 has a very clear and large LCD with audible and visible alarm and can detect accurate ozone level and send alarm to the operator in time for prevention even in very inimical work environment 2 T...

Page 4: ...battery voltage TWA STEL Ex mark Ex ibdIICT4 Protection degree IP45 Work temperature 10 50 Work humidity 5 90 RH Size 126mm L x 66mm W x 32mm H Weight 220g with charger 3 Operating instruction Menu K...

Page 5: ...her with the symbol for battery undervoltage When the power is exhausted the symbol for detection failure will flicker together with the symbol for battery undervoltage and the screen will display an...

Page 6: ...cuit and the sensor s 2 Displaying the version of software and detection range 3 Displaying Grade I alarm point combustible gas noxious gas and high concentration O2 alarm and Grade II alarm point com...

Page 7: ...ation of the gas and output a measuring result The detection principle of diffusion mode applies the above way to detecting the concentration of the surrounding gas with an advantage of swift response...

Page 8: ...alarm symbol Grade I high concentration Grade II low concentration STEL and TWA will indicates as per the gas type and alarm grade If the concentration of the gas detected is beyond the measuring rang...

Page 9: ...alarm indicating mode Mode conversion for alarm or non alarm Setting value for Grade I alarm point Setting value for Grade II alarm point Setting value for STEL alarm point Setting value for TWA alarm...

Page 10: ...ration Alarm Indication at exceeding YES means the detector works with measuring and Vibration Alarm Indication the modification can be affirmed by or unchanged if not pressing Alarm Mode Select by pr...

Page 11: ...te it usually The detector can be calibrated according to the instruction in this Section after it has been used for a long time or it is necessary to take a calibrating or readjusting Calibrating sho...

Page 12: ...Zero point calibrating Sensitivity calibrating for O2 0 0 vol For noxious gas sensors the sensitivity calibrating should be carried out every 6 months The user should apply a standard gas to sensor c...

Page 13: ...3 seconds the detector will be brought into sensitivity calibrating mode with audible indication The user can adjust the gas type for calibrating by UP Key or DOWN Key and affirm by Then it will have...

Page 14: ...intenance or replace of any part should be carried out by a specially and well trained person with original spare parts from our company 5 5 Replace or recharge of battery should be carried out in a s...

Page 15: ...environment will be beyond the guarantee scope 6 3 Our company will take no responsibility for the reliability of the detector if the user has carried out any repair or replace of any part beyond the...

Reviews: