Joyfay Delicate BX80 Operating Instructions Manual Download Page 11

 

 

TWA alarm

 

 

Increase  the  value  by 

pressing  "▲"  or 

decrease  by  "▼",  the  modification  can  be 

affirmed by "

" or unchanged if not pressing 

"

".

 

STEL alarm

   

 

Increase  the  value  by  pressing  "▲"  or 

decrease  by  "▼",  the  modification  can  be 

affirmed by "

" or unchanged if not pressing 

"

".

 

Save Indication

 

 

Select by pressing "▲" or "▼"; "NO" means 

do not save changes and "YES" means save 

changes. The modification can be affirmed by 

"

" and exit setting mode or unchanged if not 

pressing "

".

 

 

4. Adjusting and Calibrating 

The  detector  has  been  well  calibrated  before  leaving  factory,  and  user  needn't  to 

calibrate  it  usually.  The  detector  can  be  calibrated  according  to  the  instruction  in  this 

Section  after  it  has  been used for  a long  time  or it  is  necessary  to  take  a calibrating  or 

readjusting. Calibrating should be carried out only by a professional or a specially trained 

user. 

4.1 Zero point calibrating (Sensitivity calibrating for O

2

, 20.9%vol) 

Zero point calibrating should be carried out in clean air every day or every time after 

an alarm. 

At normal measuring status, press and hold down "

" and "▼" for about 5 seconds, 

the detector will be brought into zero point calibrating with an audible indication; pressing 

"▲" or "▼" to select "YES" or "NO" and affirm by " ", the screen will display "CAL" and 

flickering zero point calibrating symbol, and start a 20 seconds' countdown. 

                     

 

 

 

 

 

Summary of Contents for BX80

Page 1: ...Portable O3 Ozone Tester Detector BX80 Operating instruction manual...

Page 2: ...ontent 1 General 3 2 Technical performance and parameters 3 3 Operating instruction 4 4 Adjusting and calibrating 11 5 Daily use and maintenance 14 6 Storage and quality assurance 14 7 Complete set 15...

Page 3: ...BX80 has a very clear and large LCD with audible and visible alarm and can detect accurate ozone level and send alarm to the operator in time for prevention even in very inimical work environment 2 T...

Page 4: ...battery voltage TWA STEL Ex mark Ex ibdIICT4 Protection degree IP45 Work temperature 10 50 Work humidity 5 90 RH Size 126mm L x 66mm W x 32mm H Weight 220g with charger 3 Operating instruction Menu K...

Page 5: ...her with the symbol for battery undervoltage When the power is exhausted the symbol for detection failure will flicker together with the symbol for battery undervoltage and the screen will display an...

Page 6: ...cuit and the sensor s 2 Displaying the version of software and detection range 3 Displaying Grade I alarm point combustible gas noxious gas and high concentration O2 alarm and Grade II alarm point com...

Page 7: ...ation of the gas and output a measuring result The detection principle of diffusion mode applies the above way to detecting the concentration of the surrounding gas with an advantage of swift response...

Page 8: ...alarm symbol Grade I high concentration Grade II low concentration STEL and TWA will indicates as per the gas type and alarm grade If the concentration of the gas detected is beyond the measuring rang...

Page 9: ...alarm indicating mode Mode conversion for alarm or non alarm Setting value for Grade I alarm point Setting value for Grade II alarm point Setting value for STEL alarm point Setting value for TWA alarm...

Page 10: ...ration Alarm Indication at exceeding YES means the detector works with measuring and Vibration Alarm Indication the modification can be affirmed by or unchanged if not pressing Alarm Mode Select by pr...

Page 11: ...te it usually The detector can be calibrated according to the instruction in this Section after it has been used for a long time or it is necessary to take a calibrating or readjusting Calibrating sho...

Page 12: ...Zero point calibrating Sensitivity calibrating for O2 0 0 vol For noxious gas sensors the sensitivity calibrating should be carried out every 6 months The user should apply a standard gas to sensor c...

Page 13: ...3 seconds the detector will be brought into sensitivity calibrating mode with audible indication The user can adjust the gas type for calibrating by UP Key or DOWN Key and affirm by Then it will have...

Page 14: ...intenance or replace of any part should be carried out by a specially and well trained person with original spare parts from our company 5 5 Replace or recharge of battery should be carried out in a s...

Page 15: ...environment will be beyond the guarantee scope 6 3 Our company will take no responsibility for the reliability of the detector if the user has carried out any repair or replace of any part beyond the...

Reviews: