1. General 

BX80 O3 Ozone tester is a portable O3 gas detector/alarm that can be used to test 

the  ozone  level  in  air  or  gas.  The  tester  comprises  circuit  board,  annunciator,  LCD, 

electrochemistry  sensor.  The  sensor  changes  the  ozone  concentration  into  electricity 

signal and displays on the LCD. If you set a threshold, the tester will make alarm signal to 

remind you when the ozone concentration reach the threshold. 

 

BX80 has a very clear and large LCD with audible and visible alarm, and can detect 

accurate ozone level and send alarm to the operator in time for prevention even in very 

inimical work environment. 

2. Technical performance and parameters 

 

Features 

• Smart, portable and firm

 

• Audible and visible alarm

 

• Large screen digital and character display, instantaneous value, peak value 

and mean value available

 

•  Self-checking  for  displaying,  battery,  sensor,  audible  and  visible  alarm  at 

startup 

• Regular light flashing and audible indication 
• Distinct audible alarm

 

• Low maintenance cost

 

 

 

Main technical indexes

 

Gas for detect

 

Range 

Precision

 

Least count

 

Response time

 

Ozone (O3) 

0-5-20ppm 

±5% (F.S) 

0.01/0.1ppm 

≤ 60 sec 

Summary of Contents for BX80

Page 1: ...Portable O3 Ozone Tester Detector BX80 Operating instruction manual...

Page 2: ...ontent 1 General 3 2 Technical performance and parameters 3 3 Operating instruction 4 4 Adjusting and calibrating 11 5 Daily use and maintenance 14 6 Storage and quality assurance 14 7 Complete set 15...

Page 3: ...BX80 has a very clear and large LCD with audible and visible alarm and can detect accurate ozone level and send alarm to the operator in time for prevention even in very inimical work environment 2 T...

Page 4: ...battery voltage TWA STEL Ex mark Ex ibdIICT4 Protection degree IP45 Work temperature 10 50 Work humidity 5 90 RH Size 126mm L x 66mm W x 32mm H Weight 220g with charger 3 Operating instruction Menu K...

Page 5: ...her with the symbol for battery undervoltage When the power is exhausted the symbol for detection failure will flicker together with the symbol for battery undervoltage and the screen will display an...

Page 6: ...cuit and the sensor s 2 Displaying the version of software and detection range 3 Displaying Grade I alarm point combustible gas noxious gas and high concentration O2 alarm and Grade II alarm point com...

Page 7: ...ation of the gas and output a measuring result The detection principle of diffusion mode applies the above way to detecting the concentration of the surrounding gas with an advantage of swift response...

Page 8: ...alarm symbol Grade I high concentration Grade II low concentration STEL and TWA will indicates as per the gas type and alarm grade If the concentration of the gas detected is beyond the measuring rang...

Page 9: ...alarm indicating mode Mode conversion for alarm or non alarm Setting value for Grade I alarm point Setting value for Grade II alarm point Setting value for STEL alarm point Setting value for TWA alarm...

Page 10: ...ration Alarm Indication at exceeding YES means the detector works with measuring and Vibration Alarm Indication the modification can be affirmed by or unchanged if not pressing Alarm Mode Select by pr...

Page 11: ...te it usually The detector can be calibrated according to the instruction in this Section after it has been used for a long time or it is necessary to take a calibrating or readjusting Calibrating sho...

Page 12: ...Zero point calibrating Sensitivity calibrating for O2 0 0 vol For noxious gas sensors the sensitivity calibrating should be carried out every 6 months The user should apply a standard gas to sensor c...

Page 13: ...3 seconds the detector will be brought into sensitivity calibrating mode with audible indication The user can adjust the gas type for calibrating by UP Key or DOWN Key and affirm by Then it will have...

Page 14: ...intenance or replace of any part should be carried out by a specially and well trained person with original spare parts from our company 5 5 Replace or recharge of battery should be carried out in a s...

Page 15: ...environment will be beyond the guarantee scope 6 3 Our company will take no responsibility for the reliability of the detector if the user has carried out any repair or replace of any part beyond the...

Reviews: