If you experience discomfort during a measurement, such as pain in the wrist or other
complaints, press the START/STOP button to release the air immediately from the cuff.
Loosen the cuff and remove it from your wrist.
If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate.
Should the cuff not deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff
from the wrist and press the START/STOP button to stop inflation.
Before use, make sure the device functions safely and is in proper working condition.
Check the device, do not use the device if it is damaged in any way. The continuous use of
a damaged unit may cause injury, improper results, or serious danger.
Do not wash the cuff in a washing machine or dishwasher!
The service life of the cuff may vary by the frequency of washing, skin condition, and
storage state.
The typical service life is 10000 times.
It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of
the cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to
the local guidelines.
Manufacturer will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions, etc., to assist to service personnel in parts repair.
The operator shall not touch output of batteries and the patient simultaneously.